4.1  External Trigger Input (TRGI)
 CAUTION
Only apply signals that meet the following specifications. Signals that do not meet the 
specifications may damage the instrument because of factors such as excessive voltage.
French
 ATTENTION
Appliquer uniquement des signaux respectant les spécifications ci-dessous. Les signaux ne 
respectant pas les spécifications risquent d’endommager l'instrument, à cause de facteurs tels 
que la surtension.
External Trigger Input Terminal (TRGI)
 
TRGI: External trigger input
This terminal is used when an external signal is used as the trigger source.
Item Specifications
Input level TTL (0 to 5 V)
Minimum pulse width 1 μs
Logic Rising and falling edges
Trigger delay time Within 1 μs + 1 sample period
Externally synchronized operation Possible (through the connection of TRGI and TRGO of two DL350s)
Circuit Diagram and Timing Chart for External Trigger Input
 
Equivalent to 
LCX14
100 Ω
4.7 kΩ
100 pF