2.18  Connecting Wires to the 16-CH Voltage Input Module
What to Prepare
Wiring
  Remove approximately 7 mm of the insulation from the ends of the wires.
  Electrical wire:  0.20 mm
2
 to 1.00 mm
2
 recommended (solid wire or thin stranded wire).  
AWG size: 24-18.
   
Flat-blade Screwdriver
    Tip size:   0.4 mm (thickness) × 2.5 mm (width)
    Shape:    DIN5264-A (Use a straight driver.)
Connecting Wires to the Terminal Block
First, turn off the instrument. Next, make sure that the other end of the wire that you are connecting to 
the terminal block is not connected to the device under measurement, or make sure that the device 
under measurement that you are going to connect to is turned off.
1.
  Insert the flat-blade screwdriver into the screwdriver insertion hole right next to the wire 
insertion hole.
Insert the flat-blade screwdriver into the hole firmly until the wire insertion hole will open.
2.
  Insert the wire that you prepared into the wire insertion hole.
Insert the wire until it reaches the back of the wire insertion hole.
3.
  Pull the flat-blade screwdriver out of the screwdriver insertion hole.
When you pull out the flat-blade screwdriver, hold the wire in place so that it does not come out with the 
screwdriver.
4.
  Pull lightly on the wire to make sure that it doesn’t come out.
When you finish connecting the wire, turn on the instrument and the device under measurement.
 
Wire
Flat-blade 
Screwdriver
Terminal 
Note
Removing and Attaching the Terminal Block
It is normally not necessary to remove the terminal block, but it is possible to do so.
 Removal
  Push terminal block release levers 1 and 2 to their release positions, and pull out the terminal block.
 Attachment
  Push terminal block release levers 1 and 2 to their lock positions, and insert the terminal block. Push the 
terminal block firmly all the way to the back. However, the terminal block will not go all the way to the back 
if it is upside down. Do not try to force the terminal block all the way to the back.