4-74
IM DL350-17EN
:CHANnel<x>[:VOLTage]:INVert
Function Sets or queries whether the display is inverted
(ON) or not (OFF) for a module that can measure
analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:
INVert {<Boolean>}
:CHANnel<x>[:VOLTage]:INVert?
<x> = 1 to 4
Example
:CHANNEL1:VOLTAGE:INVERT ON
:CHANNEL1:VOLTAGE:INVERT?
-> :CHANNEL1:VOLTAGE:INVERT 1
:CHANnel<x>[:VOLTage]:LSCale?
Function Queries all the linear scaling settings of a module
that can measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale?
:CHANnel<x>[:VOLTage]:LSCale:AVALue
Function Sets or queries the scaling coefficient A that is
used during linear scaling for a module that can
measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale:
AVALue {<NRf>}
:CHANnel<x>[:VOLTage]:LSCale:AVALue?
<x> = 1 to 4
<NRf>=−9.9999E+30to−1E−30,
1E−30to9.9999E+30
Example
:CHANNEL1:VOLTAGE:LSCALE:
AVALUE 25.0000E+00
:CHANNEL1:VOLTAGE:LSCALE:AVALUE?
-> :CHANNEL1:VOLTAGE:LSCALE:
AVALUE 25.0000E+00
Description The coefficient cannot be set to 0.
:CHANnel<x>[:VOLTage]:LSCale:BVALue
Function Sets or queries the offset value B that is used
during linear scaling for a module that can
measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale:
BVALue {<NRf>}
:CHANnel<x>[:VOLTage]:LSCale:BVALue?
<x> = 1 to 4
<NRf>=−9.9999E+30to−1E−30,0,
1E−30to9.9999E+30
Example
:CHANNEL1:VOLTAGE:LSCALE:
BVALUE -25.0000E+00
:CHANNEL1:VOLTAGE:LSCALE:BVALUE?
-> :CHANNEL1:VOLTAGE:LSCALE:
BVALUE 25.0000E+00
:CHANnel<x>[:VOLTage]:LSCale:DISPlay
type?
Function Queries all the linear-scaling result display-mode
settings of a module that can measure analog
voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale:
DISPlaytype?
:CHANnel<x>[:VOLTage]:LSCale:DISPlay
type:DECimalnum
Function Sets or queries the decimal place that is used
when the linear-scaling result display mode is set
to Float for a module that can measure analog
voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale:
DISPlaytype:DECimalnum {<NRf>|AUTO}
:CHANnel<x>[:VOLTage]:LSCale:
DISPlaytype:DECimalnum?
<x> = 1 to 4
<NRf> = 0 to 3
Example
:CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:DECIMALNUM AUTO
:CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:DECIMALNUM?
-> :CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:DECIMALNUM AUTO
:CHANnel<x>[:VOLTage]:LSCale:DISPlay
type:MODE
Function Sets or queries the linear-scaling result display
mode of a module that can measure analog
voltage.
Syntax
:CHANnel<x>[:VOLTage]:LSCale:
DISPlaytype:MODE {EXPonent|FLOating}
:CHANnel<x>[:VOLTage]:LSCale:
DISPlaytype:MODE?
<x> = 1 to 4
Example
:CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:MODE EXPONENT
:CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:MODE?
-> :CHANNEL1:VOLTAGE:LSCALE:
DISPLAYTYPE:MODE EXPONENT
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals