4-76
IM DL350-17EN
:CHANnel<x1>[:VOLTage]:SCHannel<x2>?
Function Queries all settings of a 4-CH module, 16-CH
voltage input module, or 16-CH temperature/
voltage input module. If the coupling is set to TC
on a 16-CH temperature/voltage input module, a
query is made for only the label and coupling.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>?
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
BWIDth
Function Sets or queries the bandwidth limit of a 4-CH
module or 16-CH voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
BWIDth {FULL|<Frequency>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
BWIDth?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
COUPling
Function Sets or queries the input coupling setting of a
4-CH module, 16-CH voltage input module, 16-
CH temperature/voltage input module.
Syntax 4-CH module (720254)
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
COUPling {AC|DC|GND}
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
COUPling {DC|GND|OFF}
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
COUPling {DC|TC|GND|OFF}
<x1> = 1, 3
<x2> = 1 to 16
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
COUPling?
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DISPlay
Function Sets or queries whether the 4-CH module is
displayed.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DISPlay {<Boolean>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DISPlay?
<x1> = 1 to 4
<x2> = 1, 2
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel?
Function Queries all DC offset cancellation settings of a
4-CH module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel?
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel:EXECute
Function Executes DC offset canceling on a 4-CH module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel:EXECute
<x1> = 1 to 4
<x2> = 1, 2
Description This command is valid when a 4-CH module
(720254) is installed, the utility’s DC offset
cancellation and gain adjustment settings are set
to ON, and the input coupling is set to DC.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel:MODE
Function Sets or queries the on/off status of DC offset
canceling of 4-CH modules.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel:MODE {<Boolean>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
DOCancel:MODE?
<x1> = 1 to 4
<x2> = 1, 2
Description This command is valid when a 4-CH module
(720254) is installed, the utility’s DC offset
cancellation and gain adjustment settings are set
to ON, and the input coupling is set to DC.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
INVert
Function Sets or queries whether the display is inverted
(ON) or not (OFF) a 4-CH module, 16-CH voltage
input module, 16-CH temperature/voltage input
module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
INVert {<Boolean>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
INVert?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals