4-77
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LABel
Function Sets or queries the label string of the specified
sub channel for a 4-CH module, 16-CH voltage
input module, 16-CH temperature/voltage input
module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LABel {<String>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LABel?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<String> = Up to 16 characters
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale?
Function Queries all linear scaling settings of a 4-CH
module, 16-CH voltage input module, or 16-CH
temperature/voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue
Function Sets or queries the scaling coefficient A that is
used during linear scaling for a 4-CH module, 16-
CH voltage input module, 16-CH temperature/
voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<NRf>=−9.9999E+30to−1E−30,
1E−30to9.9999E+30
Example
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
AVALUE 25
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
AVALUE?
-> :CHANNEL1:VOLTAGE:SCHANNEL1:
LSCALE:AVALUE 25.0000E+00
Description The coefficient cannot be set to 0.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue
Function Sets or queries the offset value B that is used
during linear scaling for a 4-CH module, 16-CH
voltage input module, 16-CH temperature/voltage
input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<NRf>=−9.9999E+30to−1E−30,0,
1E−30to9.9999E+30
Example
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
BVALUE -25
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
BVALUE?
-> :CHANNEL1:VOLTAGE:SCHANNEL1:
LSCALE:BVALUE -25.0000E+00
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals