Commands
1
2
3
4
5
App
Index
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:DELay:REFerence:SOURce
Function Sets or queries whether to set the reference point
for measuring the delay between channels for a
waveform to a trigger point or to a waveform.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:REFerence:
SOURce {TRACe|TRIGger}
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:REFerence:SOURce?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
Example
:MEASURE:CHANNEL1:DELAY:REFERENCE:
SOURCE TRACE
:MEASURE:CHANNEL1:DELAY:REFERENCE:
SOURCE?
-> :MEASURE:CHANNEL1:DELAY:
REFERENCE:SOURCE TRACE
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:DELay:REFerenc
e:SOURce command.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:DELay:REFerence:TRACe
Function Sets or queries the reference waveform trace
used to measure the delay between channels for
a waveform.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:REFerence:
TRACe {<NRf>[,<NRf>]|MATH<x>}
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:REFerence:TRACe?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
Example
:MEASURE:CHANNEL1:DELAY:REFERENCE:
TRACE 1
:MEASURE:CHANNEL1:DELAY:REFERENCE:
TRACE?
-> :MEASURE:CHANNEL1:DELAY:
REFERENCE:TRACE 1
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:DELay:REFerenc
e:TRACe command.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:DELay:STATe
Function Sets or queries the display format of the delay
parameters of the specified channel.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:STATe {TIME|OFF|
DEGRee}
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:STATe?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
Example
:MEASURE:CHANNEL1:DELAY:STATE TIME
:MEASURE:CHANNEL1:DELAY:STATE?
-> :MEASURE:CHANNEL1:DELAY:
STATE TIME
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:DELay:STATe
command.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:DELay:VALue?
Function Queries a measured delay value of the specified
waveforms parameter.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:DELay:VALue? {<NRf>}
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
Example
:MEASURE:CHANNEL1:DELAY:VALUE?
-> :MEASURE:CHANNEL1:DELAY:
VALUE 0.0000000E+00
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:DELay:VALue?
command.
This command returns “NAN” if the value
cannot be measured.
The <NRf> is used to specify which iteration
of measurement to query after executing
statistical processing. This command returns
“NAN” if the specified value does not exist.
<NRf> can be omitted. If it is omitted, the
measured values of the newest waveform in
acquisition memory are queried. If you include
<NRf>, the measured value of the waveform
<NRf> times before the newest history
waveform will be queried.
4.22 MEASure Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals