Page #189 background imageLoading...
Page #189 background image
4-162
IM DL350-17EN
:TRIGger:OUT:TYPE
Function Sets or queries the trigger output terminal output
type.
Syntax
:TRIGger:OUT:TYPE {<NRf>}
:TRIGger:OUT:TYPE?
<NRf> = 1 to 4
1: Pulse
2: Normal
3: Sample Pulse
4: Start/Stop
Example
:TRIGGER:OUT:TYPE 2
:TRIGGER:OUT:TYPE?
-> :TRIGGER:OUT:TYPE 2
Description • Pulse
H pulses are generated for a fixed interval
when a trigger occurs.
• Normal
An L pulse is generated when a trigger occurs.
The L pulse is held during the post operation.
The instrument switches to an H pulse when
the post operation finishes.
Sample Pulse
A pulse signal is transmitted at regular intervals
when waveform acquisition is started.
• Start/Stop:
A high level signal is transmitted during
waveform acquisition, and a low level signal is
transmitted otherwise.
:TRIGger:POSition
Function Sets or queries the trigger position.
Syntax
:TRIGger:POSition {<NRf>}
:TRIGger:POSition?
<NRf> = 0 to 100 (%, in 0.1 steps)
Example
:TRIGGER:POSITION 50
:TRIGGER:POSITION?
-> :TRIGGER:POSITION 50
:TRIGger:SIMPle?
Function Queries all simple trigger settings.
Syntax
:TRIGger:SIMPle?
:TRIGger[:SIMPle]:EXTernal:SLOPe
Function Sets or queries the external trigger slope.
Syntax
:TRIGger[:SIMPle]:EXTernal:
SLOPe {FALL|RISE}
:TRIGger[:SIMPle]:EXTernal:SLOPe?
Example
:TRIGGER:SIMPLE:EXTERNAL:SLOPE RISE
:TRIGGER:SIMPLE:EXTERNAL:SLOPE?
-> :TRIGGER:SIMPLE:EXTERNAL:
SLOPE RISE
:TRIGger[:SIMPle]:HYSTeresis
Function Sets or queries the hysteresis for simple triggers.
Syntax
:TRIGger[:SIMPle]:HYSTeresis {HIGH|
LOW|MIDDle}
:TRIGger[:SIMPle]:HYSTeresis?
Example
:TRIGGER:SIMPLE:HYSTERESIS HIGH
:TRIGGER:SIMPLE:HYSTERESIS?
-> :TRIGGER:SIMPLE:HYSTERESIS HIGH
Description You cannot use this command to set the setting
when the trigger source is EXTernal or TIME.
:TRIGger[:SIMPle]:LEVel
Function Sets or queries the trigger level of the channel
specified by
:TRIGger:SIMPle:SOURce
for
simple triggers .
Syntax
:TRIGger[:SIMPle]:LEVel {<Voltage>|
<NRf>|<Current>}
:TRIGger[:SIMPle]: LEVel?
Example
:TRIGGER:SIMPLE:LEVEL 1
:TRIGGER:SIMPLE:LEVEL?
-> :TRIGGER:SIMPLE:LEVEL 1.00000E+00
Description You cannot use this command to set the setting
when the trigger source is EXTernal or TIME.
Selectable range:
The selectable range is the range that can be
measured with the current measurement range
setting.
Example: If V/Div is 10, the selectable range
is ±100 V. For strain measurements,
the selectable range is ±the current
measurement range setting.
Values do not take linear scaling information
into consideration.
The trigger’s selectable range of Au7Fe
temperatures is 0 K to 280 K (273°C to 7°C)
4.33 TRIGger Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals