4-45
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
:CHANnel<x1>:CAN:SCHannel<x2>:BCOunt
(Byte Count)
Function Sets or queries the byte count of the specified
sub channel on a module that can monitor a CAN
bus.
Syntax
:CHANnel<x1>:CAN:SCHannel<x2>:
BCOunt {AUTO|<NRf>}
:CHANnel<x1>:CAN:SCHannel<x2>:
BCOunt?
CAN Bus Monitor Module (720240)
<x1> = 1 to 4
<x2> = 1 to 60
<NRf> = 1 to 8
CAN & LIN Bus Monitor Module (720241)
<x> = 1, 3
<x2> = 1 to 60
<NRf> = 1 to 8
CAN FD Monitor Module (720242)
<x> = 1 to 4
<x2> = 1 to 60
<NRf> = 1 to 64
Example
:CHANNEL1:CAN:SCHANNEL1:BCOUNT 8
:CHANNEL1:CAN:SCHANNEL1:BCOUNT? ->
:CHANNEL1:CAN:SCHANNEL1:BCOUNT 8
:CHANnel<x1>:CAN:SCHannel<x2>:BICou
nt (Bit Count)
Function Sets or queries the bit length of the data to be
extracted of the specified sub channel on a
module that can monitor CAN bus signals.
Syntax
:CHANnel<x1>:CAN:SCHannel<x2>:
BICount {<NRf>}
:CHANnel<x1>:CAN:SCHannel<x2>:
BICount?
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x1> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x1> = 1, 3
<x2> = 1 to 60
<NRf> = 1 to 32
Example
:CHANNEL3:CAN:SCHANNEL1:BICOUNT 10
:CHANNEL3:CAN:SCHANNEL1:BICOUNT?
-> :CHANNEL3:CAN:SCHANNEL1:
BICOUNT 10
:CHANnel<x1>:CAN:SCHannel<x2>:BLABel
<x3>
Function Sets or queries the specified bit label when the
data type of the specified sub channel on a
module that can monitor CAN bus signals is logic.
Syntax
:CHANnel<x1>:CAN:SCHannel<x2>:
BLABel<x3> {<String>}
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x1> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x1> = 1, 3
<x2> = 1 to 60
<x3> = 1 to 8
<String> = Up to 16 characters
Example
:CHANNEL3:CAN:SCHANNEL1:
BLABEL1 "AAA"
:CHANNEL3:CAN:SCHANNEL1:BLABEL1?
-> :CHANNEL3:CAN:SCHANNEL1:
BLABEL1 "AAA"
Description This command is valid when the data type
(:CHANnel<x1>:CAN:SCHannel<x2>:VTYPe) is
set to LOGic.
:CHANnel<x1>:CAN:SCHannel<x2>:BORDer
(Byte Order)
Function Sets or queries the method (endian) to use to
store in the internal memory the data of the
specified sub channel on a module that can
monitor CAN bus signals.
Syntax
:CHANnel<x1>:CAN:SCHannel<x2>:
BORDer {BIG|LITTle}
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x1> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x1> = 1, 3
<x2> = 1 to 60
Example
:CHANNEL3:CAN:SCHANNEL1:BORDER BIG
:CHANNEL3:CAN:SCHANNEL1:BORDER?
-> :CHANNEL3:CAN:SCHANNEL1:
BORDER BIG
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals