4-49
IM DL350-17EN
Commands
4
:CHANnel<x>:FREQ?
Function Queries all frequency module settings.
Syntax
:CHANnel<x>:FREQ?
<x> = 1 to 4
:CHANnel<x>:FREQ:INPut?
Function Queries all frequency module input settings.
<x> = 1 to 4
:CHANnel<x>:FREQ:INPut:BWIDth
Function Sets or queries the input signal bandwidth limit of
a frequency module.
Syntax
:CHANnel<x>:FREQ:INPut:BWIDth {FULL|
<Frequency>}
:CHANnel<x>:FREQ:INPut:BWIDth?
<x> = 1 to 4
<Frequency> = 100Hz, 1kHz, 10kHz, 100kHz
Example
:CHANNEL1:FREQ:INPUT:BWIDTH FULL
:CHANNEL1:FREQ:INPUT:BWIDTH?
-> :CHANNEL1:FREQ:INPUT:BWIDTH FULL
Description If Preset is set to AC200V or AC100V, you cannot
select FULL.
:CHANnel<x>:FREQ:INPut:CELimination
Function Sets or queries the input signal chattering
elimination of a frequency module.
Syntax
:CHANnel<x>:FREQ:INPut:
CELimination {<Time>}
:CHANnel<x>:FREQ:INPut:CELimination?
<x> = 1 to 4
<Time> = 0 to 1000 ms
Example
:CHANNEL1:FREQ:INPUT:
CELIMINATION 0.000
:CHANNEL1:FREQ:INPUT:CELIMINATION?
-> :CHANNEL1:FREQ:INPUT:
CELIMINATION 0.000
:CHANnel<x>:FREQ:INPut:COUPling
Function Sets or queries the input coupling setting of a
frequency module.
Syntax
:CHANnel<x>:FREQ:INPut:COUPling {AC|
DC}
:CHANnel<x>:FREQ:INPut:COUPling?
<x> = 1 to 4
Example
:CHANNEL1:FREQ:INPUT:COUPLING DC
:CHANNEL1:FREQ:INPUT:COUPLING?
-> :CHANNEL1:FREQ:INPUT:COUPLING DC
:CHANnel<x>:FREQ:INPut:HYSTeresis
Function Sets or queries the hysteresis of a frequency
module.
Syntax
:CHANnel<x>:FREQ:INPut:
HYSTeresis {HIGH|LOW|MIDDle}
:CHANnel<x>:FREQ:INPut:HYSTeresis?
<x> = 1 to 4
Example
:CHANNEL1:FREQ:INPUT:HYSTERESIS LOW
:CHANNEL1:FREQ:INPUT:HYSTERESIS?
-> :CHANNEL1:FREQ:INPUT:
HYSTERESIS LOW
:CHANnel<x>:FREQ:INPut:PRESet
Function Sets or queries the preset setting of a frequency
module.
Syntax
:CHANnel<x>:FREQ:INPut:
PRESet {AC100v|AC200v|EMPickup|
LOG12v|LOG24v|LOG3v|LOG5v|PULLup|
USER|ZERO}
:CHANnel<x>:FREQ:INPut:PRESet?
<x> = 1 to 4
Example
:CHANNEL1:FREQ:INPUT:PRESET USER
:CHANNEL1:FREQ:INPUT:PRESET?
-> :CHANNEL1:FREQ:INPUT:PRESET USER
:CHANnel<x>:FREQ:INPut:PROBe
Function Sets or queries the probe attenuation of a
frequency module.
Syntax
:CHANnel<x>:FREQ:INPut:PROBe {<NRf>}
:CHANnel<x>:FREQ:INPut:PROBe?
<x> = 1 to 4
<NRf> = 1, 10
Example
:CHANNEL1:FREQ:INPUT:PROBE 1
:CHANNEL1:FREQ:INPUT:PROBE?
-> :CHANNEL1:FREQ:INPUT:PROBE 1
:CHANnel<x>:FREQ:INPut:PULLup
Function Sets or queries whether pull-up is turned on for a
frequency module.
Syntax
:CHANnel<x>:FREQ:INPut:
PULLup {<Boolean>}
:CHANnel<x>:FREQ:INPut:PULLup?
<x> = 1 to 4
Example
:CHANNEL1:FREQ:INPUT:PULLUP ON
:CHANNEL1:FREQ:INPUT:PULLUP?
-> :CHANNEL1:FREQ:INPUT:PULLUP 1
4.6 CHANnel Group