4-58
IM DL350-17EN
:CHANnel<x1>:LIN:SCHannel<x2>:SBIT
(Start Bit)
Function Sets or queries the bit number of the extraction
start position of the specified sub channel on a
module that can monitor LIN bus signals.
Syntax
:CHANnel<x1>:LIN:SCHannel<x2>:
SBIT {<NRf>}
:CHANnel<x1>:LIN:SCHannel<x2>:SBIT?
<x1> = 2, 4
<x2> = 1 to 60
<NRf> = 0 to 63
Example
:CHANNEL4:LIN:SCHANNEL1:SBIT 0
:CHANNEL4:LIN:SCHANNEL1:SBIT?
-> :CHANNEL4:LIN:SCHANNEL1:SBIT 0
:CHANnel<x1>:LIN:SCHannel<x2>:SCALe
Function Sets or queries the scale using upper and lower
limits of the specified sub channel on a module
that can monitor LIN bus signals.
Syntax
:CHANnel<x1>:LIN:SCHannel<x2>:
SCALe {AUTO|DEFault|<NRf>,<NRf>}
:CHANnel<x1>:LIN:SCHannel<x2>:SCALe?
<x1> = 2, 4
<x2> = 1 to 60
<NRf> = –10.000E+30 to 10.000E+30
Example
:CHANNEL4:LIN:SCHANNEL1:
SCALE -1.0,1.0
:CHANNEL4:LIN:SCHANNEL1:SCALE?
-> :CHANNEL4:LIN:SCHANNEL1:
SCALE -1.000E+00,+1.000E+00
Description This command is valid when the data type
(:CHANnel<x1>:LIN:SCHannel<x2>:VTYPe) is
set to UNSigned or SIGNed.
:CHANnel<x1>:LIN:SCHannel<x2>:UNIT
Function Sets or queries the data unit of the specified sub
channel on a module that can monitor LIN bus
signals.
Syntax
:CHANnel<x1>:LIN:SCHannel<x2>:
UNIT {<String>}
:CHANnel<x1>:LIN:SCHannel<x2>:UNIT?
<x1> = 2, 4
<x2> = 1 to 60
<String> = Up to 16 characters
Example
:CHANNEL4:LIN:SCHANNEL1:UNIT “UNIT”
:CHANNEL4:LIN:SCHANNEL1:UNIT?
-> :CHANNEL4:LIN:SCHANNEL1:
UNIT “UNIT”
Description This command is valid when the data type
(:CHANnel<x1>:LIN:SCHannel<x2>:VTYPe) is
set to UNSigned or SIGNed.
:CHANnel<x1>:LIN:SCHannel<x2>:VTYPe
(Value Type)
Function Sets or queries the data type of the specified sub
channel on a module that can monitor LIN bus
signals.
Syntax
:CHANnel<x1>:LIN:SCHannel<x2>:
VTYPe {UNSigned|SIGNed|LOGic}
:CHANnel<x1>:LIN:SCHannel<x2>:VTYPe?
<x1> = 2, 4
<x2> = 1 to 60
Example
:CHANNEL4:LIN:SCHANNEL1:
VTYPE UNSIGNED
:CHANNEL4:LIN:SCHANNEL1:VTYPE?
-> :CHANNEL4:LIN:SCHANNEL1:
VTYPE UNSIGNED
:CHANnel<x1>:LIN:SCHannel<x2>:ZOOM
Function Sets or queries the vertical zoom factor of logic
waveforms on a module that can monitor LIN bus
signals.
Syntax
:CHANnel<x1>:LIN:SCHannel<x2>:
ZOOM {<NRf>}
:CHANnel<x1>:LIN:SCHannel<x2>:ZOOM?
<x1> = 2, 4
<x2> = 1 to 60
<NRf> = 0.1, 0.111, 0.125, 0.143, 0.167, 0.2, 0.25,
0.33, 0.4, 0.5, 0.556, 0.625, 0.667, 0.714,
0.8, 0.833, 1, 1.11, 1.25, 1.33, 1.43, 1.67,
2, 2.22, 2.5, 3.33
Example
:CHANNEL4:LIN:SCHANNEL1:ZOOM 1.0
:CHANNEL4:LIN:SCHANNEL1:ZOOM?
-> :CHANNEL4:LIN:SCHANNEL1:ZOOM 1.0
Description This command is valid when the data type is
logic.
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals