Index
:CHANnel<x>:SENT:ERRor:TRIGger:PNUMb
er (Pulse Number)
Function Sets or queries whether pulse number errors of
a SENT signal are to be displayed in error trigger
channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
PNUMber {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:
PNUMber?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER ON
:CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER ON
:CHANnel<x>:SENT:ERRor:TRIGger:SCPul
se (Succesive Calibration Pulse)
Function Sets or queries whether successive calibration
pulse errors of a SENT signal are to be displayed
in error trigger channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
SCPulse {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:
SCPulse?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE ON
:CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE ON
:CHANnel<x>:SENT:ERRor:TRIGger:SCRC
(Slow Channel CRC)
Function Sets or queries whether slow channel CRC errors
of a SENT signal are to be displayed in error
trigger channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
SCRC {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:SCRC?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:SCRC ON
:CHANNEL1:SENT:ERROR:TRIGGER:SCRC?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
SCRC ON
:CHANnel<x>:SENT:PROBe
Function Sets or queries the probe attenuation of a SENT
signal channel.
Syntax
:CHANnel<x>:SENT:PROBe {<NRf>}
:CHANnel<x>:SENT:PROBe?
<x> = 1 to 4
<NRf> = 1, 10
Example
:CHANNEL1:SENT:PROBe 1
:CHANNEL1:SENT:PROBe?
-> :CHANNEL1:SENT:PROBe 1
:CHANnel<x1>:SENT:SCHannel<x2>?
Function Queries all sub channel settings of a SENT
monitor module.
Syntax
:CHANnel<x>:SENT:SCHannel<x2>?
<x1> = 1 to 4
<x2> = 1 to 11
:CHANnel<x1>:SENT:SCHAnnel<x2>:BICou
nt (Bit Count)
Function Sets or queries the bit length of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
BICount {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:
BICount?
<x1> = 1 to 4
<x2> = 1 to 3, 5 to 9
<NRf>
Fast CH: 1 to 32
Slow CH (Short): 1 to 8
Slow CH (Enhanced ID 4bit+Data 16bit):
1 to 16
Slow CH (Enhanced ID 8bit+Data 12bit):
1 to 12
Example
:CHANNEL1:SENT:SCHANNEL1:BICOUNT 12
:CHANNEL1:SENT:SCHANNEL1:BICOUNT?
-> :CHANNEL1:SENT:SCHANNEL1:
BICOUNT 12
Description This command can be used on a sub channel
whose data type is set to FastCH or SlowCH.
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals