EasyManua.ls Logo

YOKOGAWA DL350 - Page 90

YOKOGAWA DL350
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-63
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
:CHANnel<x>:SENT:ERRor:TRIGger:PNUMb
er (Pulse Number)
Function Sets or queries whether pulse number errors of
a SENT signal are to be displayed in error trigger
channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
PNUMber {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:
PNUMber?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER ON
:CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
PNUMBER ON
:CHANnel<x>:SENT:ERRor:TRIGger:SCPul
se (Succesive Calibration Pulse)
Function Sets or queries whether successive calibration
pulse errors of a SENT signal are to be displayed
in error trigger channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
SCPulse {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:
SCPulse?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE ON
:CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
SCPULSE ON
:CHANnel<x>:SENT:ERRor:TRIGger:SCRC
(Slow Channel CRC)
Function Sets or queries whether slow channel CRC errors
of a SENT signal are to be displayed in error
trigger channels.
Syntax
:CHANnel<x>:SENT:ERRor:TRIGger:
SCRC {<Boolean>}
:CHANnel<x>:SENT:ERRor:TRIGger:SCRC?
<x> = 1 to 4
Example
:CHANNEL1:SENT:ERROR:TRIGGER:SCRC ON
:CHANNEL1:SENT:ERROR:TRIGGER:SCRC?
-> :CHANNEL1:SENT:ERROR:TRIGGER:
SCRC ON
:CHANnel<x>:SENT:PROBe
Function Sets or queries the probe attenuation of a SENT
signal channel.
Syntax
:CHANnel<x>:SENT:PROBe {<NRf>}
:CHANnel<x>:SENT:PROBe?
<x> = 1 to 4
<NRf> = 1, 10
Example
:CHANNEL1:SENT:PROBe 1
:CHANNEL1:SENT:PROBe?
-> :CHANNEL1:SENT:PROBe 1
:CHANnel<x1>:SENT:SCHannel<x2>?
Function Queries all sub channel settings of a SENT
monitor module.
Syntax
:CHANnel<x>:SENT:SCHannel<x2>?
<x1> = 1 to 4
<x2> = 1 to 11
:CHANnel<x1>:SENT:SCHAnnel<x2>:BICou
nt (Bit Count)
Function Sets or queries the bit length of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
BICount {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:
BICount?
<x1> = 1 to 4
<x2> = 1 to 3, 5 to 9
<NRf>
Fast CH: 1 to 32
Slow CH (Short): 1 to 8
Slow CH (Enhanced ID 4bit+Data 16bit):
1 to 16
Slow CH (Enhanced ID 8bit+Data 12bit):
1 to 12
Example
:CHANNEL1:SENT:SCHANNEL1:BICOUNT 12
:CHANNEL1:SENT:SCHANNEL1:BICOUNT?
-> :CHANNEL1:SENT:SCHANNEL1:
BICOUNT 12
Description This command can be used on a sub channel
whose data type is set to FastCH or SlowCH.
4.6 CHANnel Group

Table of Contents

Other manuals for YOKOGAWA DL350

Related product manuals