4-65
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
:CHANnel<x1>:SENT:SCHAnnel<x2>:FACT
or
Function Sets or queries the scaling coefficient (value per
bit) of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
FACTor {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:
FACTor?
<x1> = 1 to 4
<x2> = 1 to 3, 5 to 9
<NRf>=−10.000E−30to10.000E+30
Example
:CHANNEL1:SENT:SCHANNEL1:FACTOR 1.0
:CHANNEL1:SENT:SCHANNEL1:FACTOR?
-> :CHANNEL1:SENT:SCHANNEL1:
FACTOR 1.0
Description This command can be used on a sub channel
whose data type is set to FastCH or SlowCH.
:CHANnel<x1>:SENT:SCHannel<x2>:INPut
Function Sets or queries the input on/off state of SENT
data.
Syntax
:CHANnel<x>:SENT:SCHannel<x2>:
INPut {<Boolean>}
:CHANnel<x>:SENT:SCHannel<x2>:INPut?
<x1> = 1 to 4
<x2> = 1 to 11
Example
:CHANNEL1:SENT:SCHANNEL1:INPUT ON
:CHANNEL1:SENT:SCHANNEL1:INPUT?
-> :CHANNEL1:SENT:SCHANNEL1:INPUT ON
:CHANnel<x1>:SENT:SCHAnnel<x2>:LABel
Function Sets or queries the display label of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
LABel {<String>}
:CHANnel<x1>:SENT:SCHannel<x2>:
LABel?
<x1> = 1 to 4
<x2> = 1 to 11
<String> = Up to 16 characters
Example
:CHANNEL1:SENT:SCHANNEL1:
LABEL “FastCH1”
:CHANNEL1:SENT:SCHANNEL1:LABEL?
-> :CHANNEL1:SENT:SCHANNEL1:
LABEL “FastCH1”
:CHANnel<x1>:SENT:SCHAnnel<x2>:OFFS
et
Function Sets or queries the offset value of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
OFFSet {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:
OFFSet?
<x1> = 1 to 4
<x2> = 1 to 3, 5 to 9
<NRf>=−10.000E−30to10.000E+30
Example
:CHANNEL1:SENT:SCHANNEL1:OFFSET 0.0
:CHANNEL1:SENT:SCHANNEL1:OFFSET?
-> :CHANNEL1:SENT:SCHANNEL1:
OFFSET 0.0
Description This command can be used on a sub channel
whose data type is set to FastCH or SlowCH.
:CHANnel<x1>:SENT:SCHAnnel<x2>:POSit
ion
Function Sets or queries the display position of SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
POSition {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:
POSition?
<x1> = 1 to 4
<x2> = 1 to 11
<NRf> = –5.00 to 5.00
Example
:CHANNEL1:SENT:SCHANNEL1:
POSITION 0.0
:CHANNEL1:SENT:SCHANNEL1:POSITION?
-> :CHANNEL1:SENT:SCHANNEL1:
POSITION 0.0
:CHANnel<x1>:SENT:SCHAnnel<x2>:SBIT
(Start Bit)
Function Sets or queries the data extraction position of
SENT data.
Syntax
:CHANnel<x1>:SENT:SCHannel<x2>:
SBIT {<NRf>}
:CHANnel<x1>:SENT:SCHannel<x2>:SBIT?
<x1> = 1 to 4
<x2> = 1 to 3, 5 to 9
<NRf>
Fast CH: 0 to 31
Slow CH (Short): 0 to 7
Slow CH (Enhanced ID 4bit+Data 16bit):
0 to 15
Slow CH (Enhanced ID 8bit+Data 12bit):
0 to 11
Example
:CHANNEL1:SENT:SCHANNEL1:SBIT 0
:CHANNEL1:SENT:SCHANNEL1:SBIT?
-> :CHANNEL1:SENT:SCHANNEL1:SBIT 0
Description This command can be used on a sub channel
whose data type is set to FastCH or SlowCH.
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals