Commands
4
:CHANnel<x>:STRain?
Function Queries all settings of a strain module.
<x> = 1 to 4
:CHANnel<x>:STRain:BALance?
Function Queries the ON/OFF state of target channels
when executing balancing collectively on all
channels of a strain module.
Syntax
:CHANnel<x>:STRain:BALance?
Description <x> can be any number provided that it
corresponds to the number of a channel of a
strain module.
:CHANnel<x1>:STRain:BALance:CHANnel
<x2>
Function Sets or queries the channels that balancing will
be performed on for a strain module.
Syntax
:CHANnel<x1>:STRain:BALance:
CHANnel<x2> {<Boolean>}
:CHANnel<x1>:STRain:BALance:
CHANnel<x2>?
Example
:CHANNEL1:STRAIN:BALANCE:CHANNEL1 ON
:CHANNEL1:STRAIN:BALANCE:CHANNEL1?
-> :CHANNEL1:STRAIN:BALANCE:
CHANNEL1 1
Description For specifying the target channels for executing
balancing collectively on all channels, <x1, x2>: 1
to 4.
Select channels in which strain modules are
installed.
:CHANnel<x>:STRain:BALance:EXECute
Function Executes strain balancing collectively on all strain
modules.
Syntax
:CHANnel<x>:STRain:BALance:EXECute
<x> = 1 to 4
Description This command simultaneously balances channels
that are switched on with the
:CHANnel<x1>:STRain:BALance:CHANnel<x2>
command.
<x> can be any number provided that it
corresponds to the number of a channel of a
strain module.
:CHANnel<x>:STRain:BWIDth
Function Sets or queries the filter of a strain module.
Syntax
:CHANnel<x>:STRain:BWIDth {FULL|
<Frequency>}
:CHANnel<x>:STRain:BWIDth?
<x> = 1 to 4
<Frequency> = 10Hz, 100Hz, 1kHz
Example
:CHANNEL1:STRAIN:BWIDTH 10.00E+00
:CHANNEL1:STRAIN:BWIDTH?
-> :CHANNEL1:STRAIN:BWIDTH 10.00E+00
:CHANnel<x>:STRain:EXCitation
Function Sets or queries the bridge voltage of a strain
module.
Syntax
:CHANnel<x>:STRain:
EXCitation {<Voltage>}
:CHANnel<x>:STRain:EXCitation?
<x> = 1 to 4
<Voltage> = 2V, 5V, 10V
Example
:CHANNEL1:STRAIN:EXCITATION 5
:CHANNEL1:STRAIN:EXCITATION?
-> :CHANNEL1:STRAIN:EXCITATION 5
:CHANnel<x>:STRain:GFACtor
Function Sets or queries the gauge factor of a strain
module.
Syntax
:CHANnel<x>:STRain:GFACtor {<NRf>}
:CHANnel<x>:STRain:GFACtor?
<x> = 1 to 4
<NRf> = 1.90 to 2.20
Example
:CHANNEL1:STRAIN:GFACTOR 1.9
:CHANNEL1:STRAIN:GFACTOR?
-> :CHANNEL1:STRAIN:GFACTOR 1.90
:CHANnel<x>:STRain:INVert
Function Sets or queries whether the display is inverted
(ON) or not (OFF) for a strain module.
Syntax
:CHANnel<x>:STRain:
INVert {<Boolean>}
:CHANnel<x>:STRain:INVert?
<x> = 1 to 4
Example
:CHANNEL1:STRAIN:INVERT ON
:CHANNEL1:STRAIN:INVERT?
-> :CHANNEL1:STRAIN:INVERT 1
:CHANnel<x>:STRain:LSCale?
Function Queries all the linear scaling settings of a strain
module.
Syntax
:CHANnel<x>:STRain:LSCale?
:CHANnel<x>:STRain:LSCale:AVALue
Function Sets or queries the scaling coefficient A that is
used during linear scaling for a strain module.
Syntax
:CHANnel<x>:STRain:LSCale:
AVALue {<NRf>}
:CHANnel<x>:STRain:LSCale:AVALue?
<x> = 1 to 4
<NRf>=−9.9999E+30to−1E−30,
1E−30to9.9999E+30
Example
:CHANNEL1:STRAIN:LSCALE:
AVALUE 25.0000E+00
:CHANNEL1:STRAIN:LSCALE:AVALUE?
-> :CHANNEL1:STRAIN:LSCALE:
AVALUE 25.0000E+00
Description The coefficient cannot be set to 0.
4.6 CHANnel Group