Page #99 background imageLoading...
Page #99 background image
4-72
IM DL350-17EN
:CHANnel<x1>:TEMPerature:SCHannel<x2
>:SCALe
Function Sets or queries the upper and lower limits on the
screen when temperature measurements are
performed on a 16-CH temperature/voltage input
module.
Syntax
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:SCALe {<NRf>,<NRf>}
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:SCALe?
<x1> = 1, 3
<x2> = 1 to 16
<NRf>=−5432to5432(in0.1steps)
Example
:CHANNEL1:TEMPERATURE:SCHANNEL1:
SCALE 1300.0,-200.0
:CHANNEL1:TEMPERATURE:SCHANNEL1:
SCALE?
-> :CHANNEL1:TEMPERATURE:SCHANNEL1:
SCALE 1300.0,-200.0
Description This command is valid when the input coupling
is set to TC. If the input coupling is not set to TC,
Error 113, “Undefined header,” will occur.
:CHANnel<x1>:TEMPerature:SCHannel<x2
>:TYPE
Function Sets or queries the thermocouple type to use
when temperature measurements are performed
on a 16-CH temperature/voltage input module.
Syntax
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:TYPE {K|E|J|T|L|U|N|R|
S|B|W|Au7fe}
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:TYPE?
<x1> = 1, 3
<x2> = 1 to 16
Example
:CHANNEL1:TEMPERATURE:SCHANNEL1:
TYPE K
:CHANNEL1:TEMPERATURE:SCHANNEL1:
TYPE?
-> :CHANNEL1:TEMPERATURE:SCHANNEL1:
TYPE K
Description This command is valid when the input coupling
is set to TC. If the input coupling is not set to TC,
Error 113, “Undefined header,” will occur.
:CHANnel<x1>:TEMPerature:SCHannel<x2
>:UNIT
Function Sets or queries the unit when temperature
measurements are performed on a 16-CH
temperature/voltage input module.
Syntax
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:UNIT {C|F|K}
:CHANnel<x1>:TEMPerature:
SCHannel<x2>:UNIT?
<x1> = 1, 3
<x2> = 1 to 16
Example
:CHANNEL1:TEMPERATURE:SCHANNEL1:
UNIT C
:CHANNEL1:TEMPERATURE:SCHANNEL1:
UNIT?
-> :CHANNEL1:TEMPERATURE:SCHANNEL1:
UNIT C
Description This command is valid when the input coupling
is set to TC. If the input coupling is not set to TC,
Error 113, “Undefined header,” will occur.
:CHANnel<x>:TEMPerature:TYPE
Function Sets or queries the thermocouple type that is
used when temperature measurements are
performed.
Syntax
:CHANnel<x>:TEMPerature:TYPE {K|E|J|
T|L|U|N|R|S|B|W|Au7fe}
:CHANnel<x>:TEMPerature:TYPE?
<x> = 1 to 4
Example
:CHANNEL1:TEMPERATURE:TYPE K
:CHANNEL1:TEMPERATURE:TYPE?
-> :CHANNEL1:TEMPERATURE:TYPE K
:CHANnel<x>:TEMPerature:UNIT
Function Sets or queries the unit that is used when
temperature measurements are performed.
Syntax
:CHANnel<x>:TEMPerature:UNIT {C|F|K}
:CHANnel<x>:TEMPerature:UNIT?
<x> = 1 to 4
Example
:CHANNEL1:TEMPERATURE:UNIT C
:CHANNEL1:TEMPERATURE:UNIT?
-> :CHANNEL1:TEMPERATURE:UNIT C
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals