IM 01C25T01-01E
3-24
3. OPERATION
(4) Software Seal
When you lose the password that has been registered, it
is possible to release the Write Protect mode by
using the general use password: “YOKOGAWA.”
When the password is used, the status shown in the
parameter of Software seal is changed from “KEEP”
to “BREAK.” The status returns to “KEEP” by
entering a newly set password at Enable wrt 10min.
3.5.14 Signal Characterizer
This function is used to compensate the output for non-
linear applications. The characterized values are
applied to the 4-20 mA output. For the measured
pressure, a maximum of nine coordinates can be
specified between 0-100%. Perform the coordinate
settings while the S.C. at S.C. menu parameter is
Disabled.
To apply the settings to the output, set the S.C.
parameter to Enabled.
Note that the EJX rejects the activation of the function
by AL. 60 with the following transmitter’s status:
• When the specified coordinates of x and y are not
incremental as the input increases.
• When the output mode of the output signal is set as
Sq root; at the same time, the low cut mode is set to
Linear.
F0342.EPS
Y
X
100%0%
100%
INPUT
OUTPUT
Input pressure in % Characterized value
F0343.EPS
(ENTER)
EJX:
Num of points
0
5
HELP DEL ESC ENTER
Enter 5 and press ENTER (F4).
1. Device setup
4. Detailed setup
2. Signal condition
S.C. menu
2. Num of points
Example: Set the number of coordinates on the line graph to
5.
1
‘ 5 ’
F0356.EPS
EJX:
S.C. menu
1 S.C. Disabled
2 Num of points 5
3 point setting
4 X Start 0.00 %
5 Y Start 0.00 %
HELP SEND HOME ENTER
Press SEND (F2) to send the data
to the transmitter, then check to
confirm that disappears.
(SEND)
SEND
2
F0344.EPS
1. Device setup
4. Detailed setup
2. Signal condition
S.C. menu
3. Point setting
Example: Set the first coordinates (X1, Y1) as (12, 14) in %.
(ENTER)
EJX:
Select one
1 List
2 Edit all points
3 Edit one point
4 Exit
ABORT ENTER
Select Edit one point, and press
ENTER (F4).
1
(ENTER)
EJX:
Specify S.C. point to
set value
1
1
DEL ABORT ENTER
Press ENTER (F4).
2
(ENTER)
EJX:
X1
10.00 %
12
HELP DEL ESC ENTER
Enter 12 and press ENTER (F4).
3
‘ 12 ’
(ENTER)
EJX:
Y1
10.00 %
14
HELP DEL ESC ENTER
Enter 14 and press ENTER (F4).
4
‘ 14 ’
×2