IM 12D7B3-E-E
Parameter setting 5-9
Code Display Function Function detail X Y Z Default values
Parameter specific functions
01 *SC.RES Select main parameter Conductivity 0 0 Cond.
Resistivity 1
02 *4-ELEC Select 2/4-EL system 2-Electrode measurement system 0 0 2-El.
4-Electrode measurement system 1
03 *0.10xC Set cell constant Press NO to step through choice of 0.100 cm
-1
multiplying factors on the second display.
0.10xC
1.00xC 0.10xC
10.0xC
100.xC
0.01xC
Press YES to select a factor
Use >, ^, ENT keys to adjust MAIN digits 1.000
04 *AIR Zero calibration Zero calibration with dry cell connected
*START Press YES to confirm selection
*”WAIT” Press YES to start, after briefly displaying
*END “WAIT”, *END will be displayed
Press YES to return to commissioning
mode
05 *POL.CK Polarization check Polarization check off 0 1 On
Polarization check on 1
06-09 Not used