IM 12D7B3-E-E
11-9. Field Change Order
11-9-1 Changes made by software release 1.1
PH201 communication added for Japanese market
11-9-2 Changes made by software release 1.2
E20 is cleared after the programmed data was recovered
11-9-3 Changes made by software release 2.1
Communication is default set to enabled / write enabled
11-9-4 Changes made by software release 2.2
Minimal celconstant changed from 0,008cm
-1
to 0,005cm
-1
11-9-5 Changes made by software release 2.3
Default Temperature Compensation Matrix loaded,
to prevent ‘impossible’ values after a loading all parameters from DCS
11-9-6 Changes made by software release 2.4
Create possibility to disable E5/E6 diagnostics by setting the E5/E6 limits to 0 (zero)
Burn down outputsignal changed to 3.9 mA when the HART communication is enabled.
When disabled it is 3.6 mA
Fixed rare HART communication failure
11-9-7 Changes made by software release 2.5
Implementation of Burn low in combination with HART changed.
Some minor improvements in HART communication.