IM 12D7B3-E-E
Parameter setting 5-13
Code Display Function Function detail X Y Z Default values
Temperature compensation functions
20 *T.R.°C Set reference temp. Use >, ^, ENT keys to set value 25 °C
21 *T.C.1 Set temp. coef. 1 Adjust compensation factor 2.1 %
if set to TC in section 5-2-5. per °C
Set value with >, ^, ENT keys
*T.C.2 Set temp. coef. 2 Adjust compensation factor 2.1 %
if set to TC in section 5-2-5. per °C
Set value with >, ^, ENT keys
22 *MATRX Select matrix Choose matrix if set to matrix comp.
in section 5-2-5, using >, ^, ENT keys
HCl (cation) pure water (0-80 °C) 1 1 HCI
Ammonia pure water (0-80 °C) 2
Morpholine pure water (0-80 °C) 3
HCl (0-5 %, 0-60 °C) 4
NaOH (0-5 %, 0-100 °C) 5
User programmable matrix 9
23 *T1 °C (°F) Set temp. range Enter 1st (lowest) matrix temp. value
*T2.. Enter 2nd matrix temp. value
*T3.. Enter 3rd matrix temp. value
*T4.. Enter 4th matrix temp. value
*T5.. Enter 5th (highest) matrix temp. value
24 *L1xT1 Enter conductivity Value for T1
*L1xT2 values for lowest Value for T2
.... concentration
*L1xT5 Value for T5
25 *L2xT1 Concentration 2 Similar to code 24
26 *L3xT1 Concentration 3 Similar to code 24
27 *L4xT1 Concentration 4 Similar to code 24
28 *L5xT1 Concentration 5 Similar to code 24
29 Not used