IM 12D07J01-10E-E
14
3.1.3. Functional specifications (at 25°C)
Temperature element SC41, SC49 : Ni100
Temperature element SC42 : Pt1000 to IEC 751
Nominal Cell Constant SC41/SC42/SC49-SP/SV24 : 0.1 cm
-1
SC41/SC42/SC49-SP/SV34 : 0.01 cm
-1
SC41/SC42/SC49 (EP08) : 10 cm
-1
SC41/SC42/SC49-EP14 (EP18) : 1 cm
-1
SC41/SC42/SC49-EP15 (EP16) : 1 cm
-1
SC41/SC42/SC49-FP/FV : 10 cm
-1
SC41/SC42/SC49-TP/TV : 10 cm
-1
Note: The SC41/SC42/SC49 temperature sensor is designed for measurement compensation
and for indication. It is NOT designed for process temperature control.
3.1.4. Dynamic specifications
Response time temperature t
90
SC41/SC42/SC49-SP/SV24 : < 3 min.
SC41/SC42/SC49-SP/SV34 : < 1 min.
SC41/SC42/SC49-EP04 (EP08) : < 3 min.
SC41/SC42/SC49-EP14 (EP18) : < 2 min.
SC41/SC42/SC49-EP15 (EP16) : < 3 min.
SC41/SC42/SC49-FP/FV : < 1 min.
SC41/SC42/SC49-TP/TV : < 1 min.
3.1.5. Operating range
Conductivity at actual process temperature : 1 µS * C.C. – 200 mS * C.C.
See Fig. 2: Measuring range of conductivity sensors section 2.1.5
Temperature SC41/SC42/SC49-SP/SV : 0°C to 150°C (32°F to 302°F)
SC41/SC42/SC49-EP : 0°C to 110°C (32°F to 230°F)
SC41/SC42/SC49-FP/FV : 0°C to 110°C (32°F to 230°F)
SC41/SC42/SC49-TP/TV : 0°C to 110°C (32°F to 230°F)
Pressure SC41/SC42/SC49-SP/SV : 0 to 10 bar (0 to 142 PSIG)
SC41/SC42/SC49-EP : 0 to 10 bar (0 to 142 PSIG)
SC41/SC42/SC49-FP/FV : 0 to 10 bar (0 to 142 PSIG)
SC41/SC42/SC49-TP/TV : 0 to 2 bar (0 to 28 PSIG)
Cable length for
Sensors with Amphenol connector or
Variopin connector directly connected
to Flexa analyzer : max. 60 meter with WU10 or WU10 in
combination with WF10 cable and BA10
junction box
For sensors with suffix -VS combined with SA11 : Optional 3 meter WE10 cable combined with
SA11 Smart Adapter