EasyManuals Logo

YOKOGAWA SC41 User Manual

YOKOGAWA SC41
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
IM 12D07J01-10E-E
15
SA11 Smart Adapter : Directly connected to the analyzer using a
WU11 cable up to 100 meters
or
Connected to a BA11 connetion box
using WU11 cable upto 100 m. The BA11
connection box is connected to the analyzer
using a WU11 cable up to 100m
3.1.6. Regulatory standards (only for SC42 with Amphenol connector)
- ATEX : Directive 2014/34/EU
by applying:
EN 60079-0
EN 60079-11
EN 60079-26
Certificate no. : DEKRA 14ATEX0074 X
II 1 G Ex ia IIC T4… T6 Ga
- IECEx
Applying standards : IEC 60079-0
: IEC 60079-11
: IEC 60079-26
Certificate no. : IECEx DEK 14.0032X
Ex ia IIC T4…T6 Ga
Conformity : EAC (Eurasia)
TS (Taiwan)
- Electrical data for ATEX/IECEx : For sensor input circuit connected to
A certified intrinsically safe circuit with the following
maximum values:
Ui = 14.4 V; Ii = 116.5 mA; Pi = 0.342 W
or
Certified intrinsically safe Yokogawa Contact
Conductivity transmitter Model FLXA21 series or
Model SC202S series.
The effective internal capacitance Ci and the
effective internal inductance Li of the sensor depends only
upon the properties and length of the integral cable.
Special conditions (X) : T6 for Tamb. -30°C to 40°C
T5 for Tamb. -30°C to 95°C
T4 for Tamb. -30°C to 130°C