IM 12D07J01-10E-E
22
3.2.2. Typical installation of SC41/SC42/SC49 sensor in FD40 Immersion fitting
The immersion fittings are for installing the SC41/SC42/SC49 sensor in tanks, open vessels or drains.
If the fitting is mounted in a tank with agitator or if it is placed in a fast flowing process, care must be
taken that the fitting is adequately supported. For this reason mounting several flanges can be ordered.
Modelcode FD40 FD40 Immersion fitting
Model code Suffix code Option code Description
FD40V28 Immersion fitting PVC
FD40S28 Immersion fitting Stainless Steel AISI 316
Immersion - Between 05 and 20 decimeter
depth (Example 06 = 6 dm. = 0.6 m.)
- NC
Flange - FN No flange
- F1 PVC flange DN50 PN10
- F2 PVC flange ANSI 2” 150 lbs
- F3 Stainless Steel flange DN50 PN10
- F4 Stainless Steel flange ANSI 2” 150 lbs
Style code * B
Protection hose /PH5 For 5.5 meter mounting cable
/PH10 For 10 meter mounting cable
Mounting cable /C05 Length 5,5 meter*
/C10 Length 10 meter*
Material Certificate /M 3.1 according to EN-10024 on SS wetted parts
Note: If it is necessary to use the fitting with another mounting cable length, this cable can be
ordered separately as WU40 model (Amphenol connector) or WU10/WE10 (Variopin
connector). This is not applicable for sensors with suffix V (Variopin model).
Model code Suffix code Description
WU10 Universal sensor
cable
Conn. type -V Variopin
Cable type -D Dual Coax
Cable length -02 2 meters
-05 5 meters
-10 10 meters
-15 15 meters
-20 20 meters
Model code Suffix code Description
WU40 Sensor cable
Cable length - LH01 1 meter
- LH02 2 meters
- LH05 5½ meters
- LH10 10 meters
- LH15 15 meters
- LH20 20 meters
- LH25 25 meters