M Series Service Manual
9650-0450-01 Rev. U 1-1
Chapter 1
Maintenance Tests
Overview
The M Series has two checkout procedures: the operators shift checklist and the extensive six-month maintenance tests checkout procedures.
Because the M Series units must be maintained ready for immediate use, it is important for users to conduct the Operators Shift Checklist
procedure at the beginning of every shift. This procedure can be completed in a few minutes and requires no additional test equipment. (See the
ZOLL M
Series Operators Guide for the Operators Shift Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to ensure that the functions of the
M
Series unit work properly. This chapter describes the step by step procedures for performing the six month maintenance test checkout. Use the
checklist at the back of this document (ZOLL M
Series Maintenance Tests Checklist) to record your results of the maintenance tests.
This chapter describes the following M Series maintenance tests:
1. Physical Inspection of the Unit
2. Front Panel Button Test
3. 3, 5, and 12 Leads Test
4. Power Supply Test (Optional)
5. Leakage Current Test
6. Paddles Test (If applicable)
7. Heart Rate Display Test
8. Calibrating Pulses on Strip Chart Test
9. Notch Filter Test
10. Heart Rate Alarm Test
11. Defibrillator Self Test
12. Synchronized Cardioversion Test

Table of Contents

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules

Related product manuals