M Series Service Manual
5-16 9650-0450-01 Rev. U
Biphasic High Voltage Module Removal
Note: For monophasic units, skip to page 5-22
1. Remove tape, unplug 2 spade connectors, and remove ground
screw from AC charger frame. Then lift ECG connector and
cable straight out.
2. Using a pair of needle nose pliers, squeeze the locking tabs
and push the MFC connector through the connector panel.
3. Disconnect Molex connector. 4. Remove printer isolator by lifting straight out.