Page 19 of 28
P/N 595369-01 Rev B 07/19
CAUTION
THE RESPIRATOR SYSTEM MUST NOT BE
PRESSURIZED WHEN BATTERIES ARE BEING
INSTALLED. DAMAGE TO THE ELECTRONIC
COMPONENTS MAY RESULT IF BATTERIES
ARE INSTALLED WHEN THE SYSTEM IS
PRESSURIZED.
A 3M Scott Air-Pak X3 Pro SCBA equipped with a PASS device require six (6) “AA”
cell batteries for operation. The six (6) batteries power the Heads-Up Display, the
PASS device, and the PAK-TRACKER Transmitter.
NOTE
THE UNIT WILL RETURN TO THE DEFAULT ID NUMBER IF THE BATTERY COVER
IS OPENED AND/OR THE BATTERIES ARE REMOVED. 3M RECOMMENDS
RE-PROGRAMMING WITH THE RFID CARD IN ACCORDANCE WITH YOUR
ORGANIZATION'S PROCEDURES TO ASSURE THAT THE USER'S UNIQUE ID
NUMBER IS IN USE AT ALL TIMES.
The batteries should be replaced only by a trained operator in a clean area known
to be nonammable.
Replace batteries as follows:
1. Close the respirator cylinder valve. Open the regulator purge valve to let out
all the trapped air. Close the regulator purge valve. Press the reset button
twice. If there is air pressure left in the system, the green light will continue to
ash while a fteen-second beep sequence sounds from the Sensor Module as
the residual air bleeds from the system. As soon as the air has completely bled
from system, the unit sounds a quick two-tone chirp and the PASS DEVICE
will be inactive. If there is no pressure in the system when the RESET button
is pressed twice, there will be no beep sequence. If there is air pressure in the
system, the PASS DEVICE will return to the active mode.
NOTE
ALWAYS BE SURE THAT CYLINDER VALVE IS OFF AND THE PASS DEVICE IS
COMPLETELY INACTIVE BEFORE CHANGING BATTERIES. NEVER REMOVE
OR REPLACE BATTERIES WHEN THE SYSTEM IS PRESSURIZED.
2. When replacing batteries on respirators, remove the cylinder and place the
respirator in a clean, non-hazardous area.
3. Use a Phillips driver to remove the Battery Housing Cover. Carefully remove
the cover and set aside.
4. Remove the used batteries by sliding them out of the battery compartment.
5. Install six (6) fresh new “AA” batteries of the same type. Always replace all
batteries at the same time.
The battery holder is marked with the style and orientation of the batteries
required.
Use six (6) each of one of the following 1.5 volt AA batteries:
Duracell
2
Alkaline MN1500
Duracell Alkaline MX1500
Duracell Alkaline PC1500
Duracell Quantum QU1500
Energizer
3
Alkaline EN91
Energizer Alkaline E91
Do not mix batteries. Verify correct orientation of batteries as shown on label
inside the battery holder.
WARNING
THE PASS DEVICE IS INTENDED TO ASSIST
IN LOCATING A PERSON WHO MAY BE IN A
LIFE-THREATENING SITUATION. FAILURE TO
FOLLOW THE INSTRUCTIONS FOR OPENING
THE BATTERY COMPARTMENT, CHANGING
THE BATTERIES, AND CLOSING THE BATTERY
COMPARTMENT MAY RESULT IN DAMAGE
THAT COULD CAUSE FAILURE OF THE PASS
DURING A LIFE-THREATENING EMERGENCY,
A FIRE, OR AN EXPLOSION IN A FLAMMABLE
OR EXPLOSIVE ATMOSPHERE, POSSIBLY
RESULTING IN INJURY OR DEATH.
WARNING
BATTERIES MUST ONLY BE CHANGED IN
AN AREA KNOWN TO BE NONFLAMMABLE.
CHANGING THE BATTERIES IN A FLAMMABLE
ATMOSPHERE MAY CAUSE AN IGNITION, WHICH
COULD RESULT IN SERIOUS INJURY OR DEATH.
BATTERY REPLACEMENT
2
Duracell is a registered trademark of The Procter & Gamble Company, Cincinnati, OH
3
Energizer is a registered trademark of Eveready Battery Company, Inc., St Louis, MO.
WARNING
TO REDUCE THE RISK OF EXPLOSION USE
BATTERIES ONLY FROM THE LIST PROVIDED,
DO NOT MIX OLD BATTERIES WITH UNUSED
BATTERIES, AND DO NOT MIX BATTERIES FROM
DIFFERENT MANUFACTURERS. UNAUTHORIZED
SUBSTITUTION OF COMPONENTS MAY IMPAIR
INTRINSIC SAFETY AND CAUSE AN EXPLOSION
WHICH COULD LEAD TO SERIOUS INJURY OR
DEATH.
USE SCREWDRIVER
TO LOOSEN SCREW
AND REMOVE COVER
WARNING
ALWAYS INSTALL THE BATTERIES IN THE
ORIENTATION SHOWN ON THE LABEL. FAILURE
TO PROPERLY INSTALL THE BATTERIES
WILL RESULT IN EITHER REDUCED OR NO
OPERATION OF THE EQUIPMENT, WHICH
COULD LEAD TO FAILURE OF THE EQUIPMENT
AND SERIOUS INJURY OR DEATH OF THE
RESPIRATOR USER.
WARNING
PROGRAMMING A PASS DEVICE USING A RFID
TAG, BADGE, OR BUTTON OVERWRITES THE
PAK-ID FOR 24 HOURS. IF THE SCBA IS USED
WITHIN 24 HOURS OF PROGRAMMING WITH
AN RFID TAG, BADGE, OR BUTTON, THE NEW
ID IS TRANSMITTED TO THE PAK-TRACKER AND
TO MONITOR. AFTER THAT TIME, THE PAK-ID
DEFAULTS TO THE ID SET USING CONFIGURE
AND THE PAK-LINK PROGRAMMER. IF NO ID
WAS SET, THEN THE ID DEFAULTS TO THE
FACTORY-SET ID.