40
Troubleshooting
ID
C
T
Short text
Full text
Corrective action
E90
E
S
Reference
Failure of reference pad
check. Reference pad
value of the tray is out of
range. See User Manual
for further instructions.
See
“13.3.2 Handling of failure
of reference pad check (E90)”
.
E89
E
S
QC lockout
Go to “QC measurement”
to perform QC check.
Perform QC check
measurements to lift the QC
lockout.
E88
E
S
Memory
limit
Database limit exceeded,
please delete results to
free up space.
Free up memory by erasing old
data.
W69
W
S
Output port
Output port not open.
Please restart the system!
Restart the device
W68
W
S
Output
internal
Output internal error.
Please restart the system!
Restart the device
W67
W
S
Output init
Output not inited. Please
restart the system!
Restart the device
W66
W
S
Output
closed
Output closed. Please
restart the system!
Restart the device
W65
W
S
Output
memory
Not enough memory for
output.Please restart the
system!
Restart the device
W64
W
S
Output
write
Cannot write output.
Please change file name or
(re)insert USB flash drive.
Use alphanumeric characters
only and make sure that the
USB flash drive is connected
properly and is detected by the
system.
If necessary, re-initialize the
USB port by tapping the 77
Elektronika logo in the top right
corner.
W63
W
S
Output
aborted
Output aborted. Please
start again.
Restart transfer.
W62
W
S
Output limit
Output reached internal
limit. Please check
protocol.
Check and verify output
settings.
W61
W
S
Output
protocol
Protocol failure. Please
check connection type.
Check and verify output
settings.
W60
W
S
Output
failure
Output failure. Please wait
and try again in a minute.
In case of repeated failure
please check connection
type.
The system continuously tries
to deliver the output. When it
can successfully display the
output, the error message will
automatically disappear.
If the error persists, check and