4544
Troubleshooting
ID
C
T
Short text
Full text
Corrective action
W131
W
A
DB failure: possible data
loss! Trying to repair. May
take some minutes, please
wait
Data was probably lost. The
system is trying to repair itself.
W131
W
P
DB failure: possible data
loss!
Check the worklist to see if
data was lost. Contact your
certified service personnel.
W130
W
P
DB is recreated. All
previous data is lost!
All existing data was lost.
Contact your certified service
personnel.
I117
I
P
Due to changes lockout
time was increased to X
day(s).
You have successfully
increased the active QC
lockout time.
I117
I
P
Successful QC check.
Lockout time was
increased to X day(s).
The QC lockout time was
restarted because of the
successful QC measurement.
I116
I
T
Reminder: Last day before
lockout.
There is only one day left
to perform a successful QC
measurement, before the QC
lockout is activated.
I115
I
A
Measure head SW update
in progress. May take
some seconds, please
wait.
N/A
I114
I
A
Connection is in progress.
Please wait.
N/A
I113
I
T
Output is paused while in
Settings » Ethernet screen.
N/A
I112
I
T
Log exported.
N/A
I111
I
T
Log export in progress.
Please wait
N/A
I110
I
T
Output paused while
navigating in settings
menu.
N/A
I109
I
T
Unused QC LOTs and
limits deleted.
N/A
I107
I
T
No password set. Please
set your password on
login!
N/A
I106
I
T
Operator added.
N/A
I105
I
T
Selection was sent for
printing
N/A
I104
I
T
Selection was sent for
output.
N/A
I103
I
T
Selection is inverted.
N/A