99
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Volumetric
Flow Units
of measure
Set units of
measure for
volumetric fl ow rate
21 VU
%FS* VU: %FS
uL/min VU: uL/min
ml/sec VU: mL/sec
ml/min VU: mL/min
ml/hr VU: mL/hr
L/sec VU: L/sec
L/min VU: L/min
L/hr VU: L/hr
L/day VU: L/day
m3/min VU: m3/min
m3/hr
VU: m3/hr
m3/day VU: m3/day
f3/sec VU: f3/sec
f3/min VU: f3/min
f3/hr VU: f3/hr
f3/day VU: f3/day
No Argument (status)
Returns currently selected units
VU: mL/min
Calibration
Settings
Sets/Reads Calibration related
parameters.
Note: Factory set Standard
conditions: 70.00 F and 14.6959
PSIA.
NOTE: Factory set normal
conditions: 32.00 F and 14.6959 PSIA
Hours since last time unit
was calibrated.
NOTE: Has to be reset to zero
each time power is removed after
calibration.
Power up pilot timer will be set
to zero each time power is
removed or instrument is reset.
22 C
S
Read/Set Units Standard
Conditions Temp [F].
Pressure [PSIA]
<Temp Value>
[F]
<Pres. Value>
[PSIA]
CS:<Tvalue>,<Pvalue
e> Example: S:70.0,14.696
No Argument
(Returns)
CS:<Tvalue>,<Pvalue
e> Example: S:70.0,14.696
T
Read/Reset device main
Calibration/Maintenance Timer
(timer values)
CT:<valve>
Example: CT:1024.2
Z
Reset Timer
CT:Z
N
Read/Set Units Normal
Conditions Temp [F].
Pressure [PSIA]
<Temp Value>
[F]
<Pres. Value>
[PSIA]
CN:<Tvalue>,<Pvalue>
Example: N:32.0.14.696
No Argument (Returns
current STP values)
CN:<Tvalue>,<Pvalue>
Example: N:32.0.14.696
P
Read Time elapsed from
Meter Power Up in hours
No Argument
CP:148.7