107
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Mass Flow
Set Point
Sets/Reads Mass Flow Set
Point Value
NOTE: New Set Point value
must not exceed 125% of
instrument full scale range.
32 SP
No Argument
(returns current settings)
SP:<value>
Example: SP:10.00
<Value> (New SP Value in
currently selected mass
ow EU)
SP:<value>
Example: SP:10.00
Local
Set Point
Options
Sets/Reads Local Set Point
Option parameters
WARNING: The "Saved Local
Setpoint" parameter is saved
in the instrument's EEPROM
memory. To prevent EEPROM
burning, avoid writing "Saved
Local Setpoint" parameters
more often than one time per
minute.
33 LS
S
Local Set Point Saved
Value
No Argument
(returns current settings)
LSS:<value>
Example: LSS:10.00
<Value>(New Value in
currently selected mass
EU)
LSS:<value>
Example: LSS:10
M
Local and Digital Set Point
Merge Mode. Argument 2:
E- Enabled
D-Disabled
<New Value>
E or D
LSM:<value>
Example: LSM:D
No Argument (Ret.
Current settings)
LSM:<value>
Example: LSM:D
A
Local Set Point Auto Start
Mode. Argument 2:
E- Enabled
D-Disabled
<New Value>
E or D
LSA:<value>
Example: LSA:D
No Argument (Ret.
Current settings)
LSA:<value>
Example: LSA:D