91
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Temperature
Alarms
Sets / reads the parameters of the
temperature alarms.
Note: High alarm value has to be
more than Low alarm value.
Device Temperature Alarm conditions:
Temp. > High Limit = H
Temp. < Low Limit = L
Low < Temp. < High = N
Alarm Settings Reply parameters:
M – mode (E/D)
Hv – High settings value
Lv – Low settings value
A – Action Delay (sec)
L – Latch mode (0-1)
P – Power Up delay (sec)
12 TA C
(set Temp Alarm
High and Low limits
parameters)
<Value>
(high limit,
Kelvin.)
[253.16 °K –
343.15 °K]
<Value>
(low limit,
Kelvin)
[253.15°K –
343.14°K]
TAC:333.25,263.15
A
(Temp Alarm action
delay in sec.)
<Value>
[0-3600]
TAA:<Value>
Example: TAA:5
E
(enable Temp
Alarm)
TA:E
D
(disable temp
alarm)*
TA:D
R (read current
status)
TAR:H (high alarm)
TAR:L (low alarm)
S
(Read current
settings)
TAS:M,Hv,Lv,A,L,P
Example:
TAS:E,333.25,263.15,2,0,10
P
(Temp Alarm Power
Up delay in sec.)
<Value>
[0-3600]
TAP:<Value>
Example: TAP:60
L
(Latch mode)
<Value>
(0-disabled*)
(1-enabled)
TAL:<Value>
where:
Value = 0 – 1
Example:
TAL:0