92
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Pressure
Alarms
Sets / reads the parameters of
the pressure alarms.
Note: High alarm value
has to be more than Low alarm
value.
Device Pressure
Alarm conditions:
Pressure
> High Limit = H
Pressure. < Low Limit = L
Low < Pressure < High = N
Alarm Settings Reply parameters:
M – mode (E/D)
Hv – High settings value
Lv – Low settings value
A – Action Delay (sec)
L – Latch mode (0-1)
P – Power Up delay (sec)
13 PA C
(set Press. Alarm
High and Low limits
parameters)
<Value>
(high limit, PSIA)
[0.1 –100.00]
PSIA
<Value>
(low limit, PSIA)
[0.0 –90.99]
PSIA
PAC:60.00,10.00
A
(Press. Alarm
action delay in sec.)
<Value>
[0-3600]
PAA:<Value>
Example: PAA:5
E
(enable Press.
Alarm)
PA:E
D
(disable Press.
Alarm)*
PA:D
R
(read current
status)
PAR:N (no alarm)
PAR:H (high alarm)
PAR:L (low alarm)
S
(Read current
settings)
PAS:M,Hv,Lv,A,L,P
Example:
PAS:E,60.00,10.0
0,2,0,5
P
(Pres. Alarm Power
Up delay in sec.)
<Value>
[0-3600]
PAP:<Value>
Example: PAP:60
L
(Latch mode)
<Value>
(0-disabled*)
(1-enabled)
PAL:<Value>
where:
Value = 0 – 1
Example:
PAL:0