AD-4405A Weighing Indicator Page 33
7. Tare
The function is used to display a net value with the container weight subtracted from
the total weight, if you place an object into a container to weigh it.
Using a serial interface such as the RS-232C, you can do this from the external
equipment.
Caution
When turning the display off with Cf02=1, 2 or 3, the tare data is cleared.
When turning the power off, the tare data is cleared.
7.1. Weighing Tare
Operation Place the tare on the weighing unit. Wait for the STABLE mark to turn on
and press the TARE key to store the tare weight. The display changes to net.
Caution When the gross value is zero or negative, tare can not be used (with Cf04=0, 1 ).
To enable tare at zero or a negative gross value, specify at Cf04.
7.2. Digital Input (Preset Tare)
Caution The input value is rounded off to the unit of division (weighing interval).
In the case of Cf02=1, 2 or 3 (power-on zero), the displayed value will be
zeroed when turning the display on.
When using a multi-interval scale, the usable input range is the first range.
In the case of Cf09=1 (To inhibit preset tare), preset tare can not be used.
Preset tare is stored as one of the Code Memory data. Refer “9. Code
Memory”.
Preset tare value can be set via the serial interface.
Step 1 Press the PT key to display the stored tare value and the triangle
mark
blinking. When the tare is cleared or is not used, the value is zero.
Step 2 Enter a new tare value using the 0 - 9 keys.
ENTER key The key to store a new tare value.
The triangle
mark is turned on and the net is displayed.
ESC key The key to return to the previous stage (weighing mode or code
memory selection mode) without changing the stored value.
7.3. Clearing Tare
When pressing the TARE key while gross is zero, tare is cleared and gross is
displayed.
When zeroing with the ZERO key, tare is cleared.