AD-4405A Weighing Indicator Page 63
16.3.3. Commands to Set Parameters
Set the Limit/Setpoint Value
Sets the limit or setpoint value of the comparison. The decimal point is not necessary.
Template Sm,n, [ value ]
m: code memory number, 0 - 4
n : setpoint order number, refer to “10.1.2. Setting the Upper/Lower
Limit values” and “10.2.3. Setting the Parameters of Setpoint
Comparison".
Command
Reply
Set the Zero Band
Sets the F26 value (zero band) of the comparison. The decimal point is not
necessary.
Template SZ, [ value ]
Command
Reply
Set the Preset Tare
Sets the preset tare value. The decimal point is not necessary.
Template PT,m, [ value ]
m: code memory number, 0 - 4
Command
Reply
Set the Unit Weight for Counting Mode
Sets the unit weight value with decimal point.
Template UW,m, [ value ]
m: code memory number, 0 - 4
Command
Reply