EasyManua.ls Logo

A&D EK-4100i - 13-2. EW-i Series

A&D EK-4100i
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
MODEL EK-410i EK-610i EK-4100i EK-6100i
Weight capacity 400 g 600 g 4000 g 6000 g
Min. display “d” 0.01 g 0.01 g 0.1 g 0.1 g
No. of samples 5, 10, 25, 50 or 100 pieces
Max. count *) 40,000 pcs 60,000 pcs 40,000 pcs 60,000 pcs
Min. unit weight *) 0.01 g 0.01 g 0.1 g 0.1 g
Min. % display 0.1 %
Min. 100 % weight 1 g 1 g 10 g 10 g
Repeatability
(Std. deviation)
0.01 g 0.01 g 0.1 g 0.1 g
Linearity ±0.02 g ±0.02 g ±0.2 g ±0.2 g
Sensitivity drift ±20 ppm / °C (10°C~30°C / 50°F~86°F)
Display 7 segment LCD display with backlight (Character height 16 mm)
Display update 10 time per second
Operating temp. -10°C~40°C / 14°F~104°F, less than 85% R.H. (non-condensing)
Power supply AC adapter or optional Ni-MH battery pack
Battery operation Approximately 9 hours (backlight off)
Weighing pan size 110 mm ø
133mm x 170mm
Weight
(approximately)
1.1 kg 1.1 kg 1.5 kg 1.5 kg
Calibration weight
(factory setting)
400 g 600 g 4000 g 6000 g
*) In case of Umin 0” (factory setting)
13-2. EW-i series
MODEL EW-150i EW-1500i EW-12Ki
Weight capacity
30 g 60 g 150 g 300 g 600 g 1500 g 3 kg 6 kg 12 kg
Min. display ”d”
0.01g 0.02g 0.05g 0.1g 0.2g 0.5g 1g 2g 5g
No. of samples 5, 10, 25, 50 or 100 pieces
Max. count *) 15,000 pcs 15,000 pcs 12,000 pcs
Min. unit weight *) 0.01 g 0.1 g 1 g
Min. % display 0.1 %
Min. 100 % weight 1 g 10 g 100 g
Repeatability
(Std. deviation)
0.01g 0.02g 0.05g 0.1g 0.2g 0.5g 1g 2g 5g
Linearity
±0.01g ±0.02g ±0.05g ±0.1g ±0.2g ±0.5g ±1g ±2g ±5g
Sensitivity drift ±20 ppm / °C (10°C~30°C / 50°F~86°F)
Display 7 segment LCD display with backlight (Character height 16 mm)
Display update 10 time per second
Operating temp. -10°C~40°C / 14°F~104°F, less than 85% R.H. (non-condensing)
Power supply AC adapter or optional Ni-MH battery pack
Battery operation Approximately 9 hours (backlight off)
Weighing pan size 110 mm ø 133 mm x 170 mm
Weight
(approximately)
1.1 kg 1.5 kg 1.5 kg
Calibration weight
(factory setting)
150 g 1500 g 12 kg
*) In case of “Umin 0” (factory setting)

Other manuals for A&D EK-4100i