Troubleshooting 9-21
6. Place the measuring tube on a clean counter with the red or
pink aperture facing up. Drop two drops of bleach onto the
aperture and let it soak for ten minutes.
7. Rinse off the bleach with distilled water.
8. Put the reference electrode and measuring tube
back together, and slide the U-shaped clip back
into place.
9. Push the probe back into the body of the dilu-
tion chamber and reattach tubing.
10. Reattach the black wire.
11. Run a blank.
12. If the blank displays flags and results in red, contact Abaxis Technical Support.