13
7 PROGRAMMING
Operating
Parameters
Secure
Parameters
20.9
% Oxygen
700
Cell Temperature
0.0
Cell milliVolts
20.00
Alarm 1 Setpoint
5.00
Alarm 2 Setpoint
Section 6.2
-----
OXYGEN CAL. SEQ.
0.0
Cell Constant mV
100.0
Span % of Theory
00000
Cal. User Code
Preset Cal.
One Point Cal.
-----
Two Point Cal.
0.0
Adjust Cell Zero
-----
Span Theory NO
YES
Section 7.1
Section 7.3
20.95
00000
Security Code
Connect to Air
20.95
Calibrating Air
-----
Calibration Pass
Failed Constant
Failed Unstable
20.95
Connect to Air
20.95
Calibrating Air
1.00
Enter Span Gas
1.00
Connect Span Gas
1.00
Calibrating Span
-----
Calibration Pass
Failed Span
Failed Constant
Failed Unstable
Section 7.4
Section 7.1
Section 7.5 Section 7.7
YES
Section A.3
Section 7.6
Section 7.2
xxxxx
English
-----
Deutsch
Français
Español
-----
SET UP OUTPUTS
-----
A1 Action EA
-----
A1 Setpoint
0.0
A2 Action EA
20.00
A2 Setpoint
-----
RTX Type 4-20
-----
Log Output NO
25.00
Retrans Span
0.0
Retrans Zero
-----
Hold Outputs NO
EB
EB
General Alarm
0-20
0-10
YES
YES
-----
ELECTRICAL CAL.
-----
Calibrate YES
xxxxx
mV Zero 1 -20mV
mV Span 1 180mV
xxxxx
mV Zero 2 10mV
xxxxx
mV Span 2 40mV
100.0
Test Retrans (%)
00000
Alter Sec. Code
00000
Alter Cal. Code
25.0
Adjust CJ Value
-----
Adjust RTX Zero
-----
Adjust RTX Span
K
-----
SET UP T/C
T/C TYPE NONE
R
S
-----
700
Preset Temp
NO
B
Fig. 7.1 Overall Programming Chart