19
7 PROGRAMMING…
Preset Temp
700
-----
SET UP T/C
-----
T/C Type    NONE
              K
SET UP OUTPUTS
              R
              S
              B
7.6 Set Up Thermocouple Page
Set Up Thermocouple
Page header
Thermocouple Type
Set the thermocouple type required: Type K, R, S, B or NONE. If thermocouple
type is set to NONE, a preset temperature must be used – see below.
Preset Temperature
If a thermocouple is not used, set the process temperature.
[600 to 1400°C in 1°C steps]
Note. This frame is not displayed unless Thermocouple Type is set to
NONE.
Press   to return to the beginning of the Oxygen Calibration Sequence.
Press   to advance to the Set Up Outputs Page.