16
…7.2 Two-point Calibration
…7 PROGRAMMING
Calibration Pass/Fail
On completion a calibration status message is displayed.
Calibration Pass: Calibration sequence successful
Failed Constant: Cell offset >±10mV
(upper display shows cell constant)
Failed Span %: Cell output <90% or >110% of slope
(upper display shows measured slope)
Failed Unstable: Cell output unstable (drifting).
Note. If sensor calibration is unsuccessful then the Cell Zero mV and Span
% of Theory parameters are unaffected. The instrument continues to operate
using parameters stored during the last successful calibration.
Press   to return to the beginning of the Oxygen Calibration Sequence.
Press   to advance to the Access to Secure Parameters Page.
Failed Unstable
-----
Calibration Pass
Failed Constant
SECURITY CODE
Continued from previous page
Failed Span