Single- and multi-stream colorimetric analyzers 13 Maintenance
IM/AZT6CR–EN Rev. J 71
4. Referring to Fig. 13.11,
a. Invert the measurement cell assembly, remove the
four screws F and remove the base plate G.
b. Remove the sample tube end face sealing cap
H
and remove seal I. Discard the seal.
c. Remove all traces of fluid from inside the sample
tube J.
5. Referring to Fig. 13.12, remove the two screws K and
remove the measurement cell top cover L.
6. Referring to Fig. 13.13, disconnect the motor wiring plug
M.
Continued…
Caution. During step 4, do not allow any fluid from the
measurement cell sample tube to come into contact with
either the skin or any metallic/electronic parts of the
analyzer. The fluid is contaminated with acid from the
reagents used in the analyzer. Wipe up all spillages
immediately.
Fig. 13.11 Removing the Measurement Cell Base Plate
F
G
I
H
J
Fig. 13.12 Removing the Measurement Cell Top Cover
Fig. 13.13 Disconnecting the Motor
L
K
M