Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers 13 Maintenance
70 IM/AZT6CR–EN Rev. J
13.3.6 Replacing the Piston Tube and Piston Assembly
1. Referring to Fig. 13.8, turn handle
A
1
/4 turn clockwise,
pull the analytical section forward and allow it to rest
against its stops.
2. Referring to Fig. 13.9:
a. Disconnect the measurement cell ribbon cable B
from the connector on the back of the measurement
cell.
b. Remove the four M4 screws C and remove the
measurement cell assembly.
3. Referring to Fig. 13.10, depress catch D and remove
cover E.
Fig. 13.8 Opening the Analytical Section
Fig. 13.9 Removing the Measurement Cell Assembly
Fig. 13.10 Removing the Measurement Cell Cover