CONTROLMASTER CM15 |  UNIVERSAL PROCESS INDICATOR,  DIN  |  CI/CM15-EN REV. A
21
Icon Number/
Message
Possible cause Suggested action

NV Error


or permanent corruption of its data.

Recalibrate If necessary.
Acknowledge error.
If problem persists contact local service organization.

NV Error
Comm Bd
Failure of non-volatile memory on
communications board or permanent corruption
of its data.
Acknowledge error.
Check communications board is correctly identified by device.
If problem persists contact local service organization.

NV Error


or permanent corruption of its data.
Acknowledge error.
Check software key functionality is enabled.
If problem persists contact local service organization.

NV Error


or permanent corruption of its data.
Acknowledge error.
Check software key functionality is enabled.
If problem persists contact local service organization.