EasyManua.ls Logo

ABB Navigator 600 Silica - Security

ABB Navigator 600 Silica
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Navigator 600 Silica
Single-stream 5 Configuration
28 IM/NAV6S/SS–EN Rev. E
5.3.4 Security
'User 1' is the System Administrator and is the only user that has access to the 'Security type' parameter. User 1 details cannot be
changed by any other user.
There are two types of security:
Basic
Enables up to four users to have access to the 'Configuration' menu. Each user has a password that can have up to four digits.
A separate password can be set to gain access to the 'Calibration & Maintenance' and 'Logging' menus. All four users use this
password.
Advanced
Enables up to twelve users to have password-protected access to any of the 'Configuration', 'Calibration & Maintenance' or
'Logging' menus. Each user can be assigned a unique 20-digit (alphanumeric) case-sensitive password. A minimum password
length can be set.
User 1 can set permissions for each user to access the 'Calibration & Maintenance' and / or 'Logging' menus. Access to the
'Configuration' menu can also be set – see Section 5.3.5, page 30.
Passwords are set by User 1. If permissions are set, users can subsequently change their own passwords.
Password expiry dates can be set.
Inactive user accounts can be disabled after a set time.
Password failure limits can be set.
A minimum password length can be set.
Note. A separate password for access to the 'Calibration & Maintenance' and 'Logging' menus is not required.

Table of Contents

Related product manuals