Navigator 600 Silica
Single-stream 13 Specification
62 IM/NAV6S/SS–EN Rev. E
13 Specification
Silica Measurement
Range
Fully user programmable 0 to 5000 ppb SiO
2,
minimum range 0 to 50 ppb
Measurement Mode
Sample stream option
Single-stream
Single-stream Performance
Measurement method
Continuous chemistry and measurement operation.
Response time
<15 mins. (90 % step change)
Typical accuracy
<±2 % of reading or ±0.5 ppb (whichever is the greater)
over the range 0 to 500 ppb
<±5 % of reading over the range 500 to 5000 ppb
Repeatability
<±2 % of reading or ±0.5 ppb (whichever is the greater)
over the range 0 to 500 ppb
<±3 % of reading over the range 500 to 5000 ppb
Solution Requirements
Number
4 reagents (2.5 l bottles)
1 standard solution (0.5 l bottle)
1 cleaning solution (0.5 l bottle)
Reagent Consumption
Continuous operation mode
2.5 l max. per 90 days
Display
Color*, passive matrix, liquid crystal display (LCD) with built-in
backlight and brightness adjustment
76800 pixel display*
*A small percentage of the display pixels may be either constantly
active or inactive. Max. percentage of inoperative pixels <0.01 %
Dedicated operator keys
Group Select / Left cursor
View Select / Right cursor
Menu key
Up / Increment key
Down / Decrement key
Enter key
Mechanical Data
Ingress protection
Dimensions
Materials of construction
Sample connections
** Not evaluated for UL or CB
Diagonal display area 144 mm (5.7 in.)
IP31** – Wet section (critical components IP66)
IP66 – Transmitter
Height 638 mm (25.1 in.) plus constant head
bracket 186 mm (7.3 in.)
Width 271 mm (10.7 in.)
Depth 182 mm (7.2 in.)
Weight 15 kg (33 lbs)
Electronics enclosure 20 % glass loaded polypropylene
Main enclosure Noryl
Lower tray 10 % glass loaded polypropylene
Door Acrylic
Inlet 6 mm (
1
/4 in.) flexible hose connection
Outlet 9 mm (
1
/3 in.) flexible hose connection