Navigator 600 Silica
Single-stream 7 Calibration & Maintenance
IM/NAV6S/SS–EN Rev. E 47
7.5 Calibration
Enables manual calibrations to be performed.
Fields Description
Secondary Calibration
Value
The value of the secondary calibration solution used (10 to 5000 ppb)
– see Appendix A.2, page 64.
Note. This is the same value used in the 'Setup Calibration / Clean' options. Changing the value here
also changes the value in the 'Setup Calibration / Clean' options.
If the analyzer does not have any active alarms or warnings that can affect a calibration, the following options are available:
Baseline (2 point, resets
zero offset)
Both a zero and a secondary calibration are performed. A new calibration factor is calculated and the
Zero Offset is reset to 0.0 ppb. A recovery period follows.
A baseline calibration must be performed when the reagents have been replaced or an annual
service has been performed.
A progress bar is displayed indicating the maximum time required for the operation.
Zero (1 point)
A zero calibration is performed and a new zero offset calculated. The recovery period follows.
A progress bar is displayed indicating the time required for the operation.
Secondary (1 point)
A secondary calibration only is performed and a new calibration factor calculated. A recovery period
follows.
A progress bar is displayed indicating the time required for the operation.
Zero & Secondary
(2 point)
Both a zero and secondary calibration are performed. A new zero offset and calibration factor are
calculated. A recovery period follows.
A progress bar is displayed indicating the time required for the operation.