Navigator 600 Silica
Single-stream 13 Specification
IM/NAV6S/SS–EN Rev. E 63
Environmental Data
Ambient Operating Temperature
5 to 45 ºC (41 to 113 ºF)
Sample Temperature
5 to 55 ºC (41 to 131 ºF)
Sample Particulate
<60 microns
<10 mgl
–1
Sample Flow Rate
>5 ml/min / <500 ml/min
Sample Pressure
Atmospheric
Storage Temperature
–20 to 75 ºC (–4 to 167 ºF)
Ambient Operating Humidity
Up to 95 % RH non-condensing
Electrical
Supply ranges
110 to 240 V max. AC 50/60 Hz ± 10 % (90 to 264 V AC, 45/65 Hz)
18 to 36 V DC 10 A power supply typical (optional)
Power consumption
75 W max. – AC
100 W max. – DC
Analog Outputs
6 isolated current outputs:
galvanically isolated (to 500 V DC) from each other and all other
circuitry
fully assignable and programmable over a 0 to 20 mA range
(up to 22 mA if required)
drives maximum 750 load
Wetted Materials
PMMA (acrylic)
PP (polypropylene)
PTFE
PP (20 % glass filled)
PEEK
NBR (nitrile)
EPDM
Santoprene
PTFE (15 % polysulphane)
NORYL
Borosilicate glass
Acrylic adhesive
Alarms / Relay Outputs
One per unit:
out-of-service alarm relay
calibration in progress alarm relay
calibration failed alarm relay
maintenance / hold alarm relay
Six per unit:
fully user-assignable and alarm relays
Rating
Connectivity / Communications
Ethernet connection
Bus communications
Profibus DP V1 (optional)
Data Handling, Storage and Display
Security
Storage
Removable Secure Digital (SD) card – maximum size 2 GB
Trend analysis
Local and remote
Data transfer
SD card or FTP
Approvals, Certification and Safety
Safety Approval
cULus
CE Mark
Covers EMC & LV Directives (including latest version EN 61010)
General Safety
EN 61010–1
Overvoltage Class 11 on inputs and outputs
Pollution category 2
EMC
Emissions & immunity
Meets requirements of IEC61326 for an Industrial Environment
DS/NAV6S-EN Rev. N
Voltage 250 V AC 30 V DC
Current 5 A AC 5 A DC
Loading (non-inductive) 1250 VA 150 W
Web server with ftp: for real time monitoring, configuration,
data file access and email capability
Multi-level security: user, configuration, calibration and
maintenance pages