Navigator 600 Silica
Single-stream Appendix B Diagnostics and Alarms
70 IM/NAV6S/SS–EN Rev. E
Solution Replacement
The analyzer is shut down while solution is replaced.
OFF ON
Scheduled
calibrations
and cleans
Scheduled Secondary Cal. Missed
The scheduled secondary calibration has been missed due to either
a manual operation being performed or an out-of-service condition.
The next calibration due date is updated automatically.
The next calibration is performed at the correct date and time
if the analyzer is in service.
Cleaning In Progress
A clean routine is being run.
A progress bar on the cleaning page indicates the time to
completion.
OFF ON
Scheduled
calibrations
and cleans
Calibration In Progress
A calibration routine is being run. A progress bar indicates the time
to completion.
1. Wait for the routine to be completed.
2. Cancel the routine and allow the analyzer to complete the
recovery routine.
ON
Scheduled
calibrations
and cleans
In Annual Service
The analyzer has been taken off-line for the annual service.
Complete the servicing tasks and press the key when finished
to reset the annual service alarm and place the analyzer on-line
(after a purge and recovery period).
OFF ON
Scheduled
calibrations
and cleans
In Manual Test Setting
The analyzer has been taken off-line while manual test settings are
checked / adjusted.
1. Complete the manual test settings routine.
2. When completed, place the analyzer back on-line by exiting
via 'Normal' mode at the 'Exit via Test Mode / Normal Mode'
prompt.
OFF ON
Scheduled
calibrations
and cleans
Icon Diagnostic message
Out-of-
Service
Relay
Hold
Relay
Cal. in
Progress
Relay
Cal.
Failed
Relay
Inhibits
Table B.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 5 of 7)