FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  135 
 
 
Flowmeter sensor in high temperature version
***
 
Lining material  Flange material  Ambient temperature range (T
amb.
)
Measuring medium temperature (T
medium
)
  Minimum Maximum
Minimum Maximum
Thick PTFE***  Steel  -10 °C (14 °F) 60 °C (140 °F)
-10 °C (14 °F) 180 °C (356 °F)
Thick PTFE***  Stainless steel  -25 °C (-13 °F) 60 °C (140 °F)
-25 °C (-13 °F) 180 °C (356 °F)
  −40 °C (-40 °F)**
PFA***  Steel  -10 °C (14 °F) 60 °C (140 °F)
-10 °C (14 °F) 180 °C (356 °F)
PFA***  Stainless steel  -25 °C (-13 °F) 60 °C (140 °F)
-25 °C (-13 °F) 180 °C (356 °F)
  −40 °C (-40 °F)**
ETFE  Steel  -10 °C (14 °F) 60 °C (140 °F)
-10 °C (14 °F) 130 °C (266 °F)
ETFE  Stainless steel  -25 °C (-13 °F) 60 °C (140 °F)
-25 °C (-13 °F) 130 °C (266 °F)
  −40 °C (−40 °F)**
*  Only for China production site 
**  For (optional) low-temperature version only 
*** Only for design level ‘A’ 
 
 
Change from one to tw o columns