140  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
 
… 13 Specification 
HygienicMaster - Temperature data 
 
The temperature range offered by the device is dependent on a 
number of different factors. 
These factors include the measuring medium temperature 
T
medium
, the ambient temperature T
amb
, operating pressure 
P
medium
, liner material and the approval for explosion protection. 
 
 
Storage temperature range 
−40 to 70 °C (−40 to 158 °F) 
 
 
Maximum permissible cleaning temperature 
 
CIP media  Liner Cleaning temperature 
Steam  PTFE, PFA 150 °C (302 °F) 
Cleaning fluid  PTFE, PFA 140 °C (284 °F) 
 
•  The maximum cleaning temperature specified applies to a 
maximum ambient temperature of 25 °C (77 °F). 
  If the ambient temperature up-scales > 25 °C (> 77 °F), then 
the temperature difference to the current temperature must 
be subtracted from the max. cleaning temperature. 
•  The specified cleaning temperature may have an effect for a 
maximum of 60 minutes. 
 
 
 
 
 
Maximum Allowable Temperature Shock 
Maximum allowable temperature shock difference in °C: Any 
Temperature gradient °C/min: Any 
 
 
Change from two to one  column 
 
 
 
 
 
Maximum ambient temperature depending on measuring medium temperature 
 
Integral mount design or remote mount design 
 
Flowmeter sensor in standard version 
Process connection  Ambient temperature range (T
amb
) Measuring medium temperature (T
medium
)
Minimum* Maximum Minimum Maximum*
Flange  -20 °C (-4 °F) 60 °C (140 °F) -25 °C (-13 °F) 100 °C (112 °F)
-20 °C (-4 °F) 40 °C (104 °F) -25 °C (-13 °F) 130 °C (266 °F)***
Variable process connections  -20 °C (-4 °F) 60 °C (140 °F) -25 °C (-13 °F) 100 °C (112 °F)
-20 °C (-4 °F) 40 °C (104 °F) -25 °C (-13 °F) 130 °C (266 °F)***
 
High temperature version – from size DN 10 (⅜ in) 
Process connection  Ambient temperature range (T
amb
) Measuring medium temperature (T
medium
)
Minimum*
Maximum Minimum Maximum
Flange  -20 °C (-4 °F)
60 °C (140 °F) -25 °C (-13 °F) 180 °C (356 °F)
*  Also available in low temperature version for ambient temperatures down to −40 °C (−40 °F) 
**  For CIP/SIP cleaning, higher temperatures are permitted for limited time periods; refer to Maximum permissible cleaning temperature on page 140. 
*** For units of nominal size DN 1 to 2 , the measuring medium temperature is limited to 120 °C (248 °F) 
 
 
Change from one to tw o columns