FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  15 
 
5  Installation 
Safety instructions 
 
 WARNING 
Risk of injury due to process conditions. 
The process conditions, for example high pressures and 
temperatures, toxic and aggressive measuring media, can 
give rise to hazards when working on the device. 
•  Before working on the device, make sure that the process 
conditions do not pose any hazards. 
•  If necessary, wear suited personal protective equipment 
when working on the device. 
•  Depressurize and empty the device / piping, allow to cool 
and purge if necessary. 
 
 
 WARNING 
Risk of injury due to live parts! 
When the housing is open, contact protection is not provided 
and EMC protection is limited. 
•  Before opening the housing, switch off the power supply. 
 
 
Use in Potentially Explosive Atmospheres 
 
 
 DANGER 
Danger of explosion if the device is operated with the 
transmitter housing or terminal box open! 
While using the device in potentially explosive atmospheres 
before opening the transmitter housing or the terminal box, 
note the following points: 
•  A valid fire permit must be present. 
•  Make sure that no flammable or hazardous atmospheres 
are present. 
 
 
Note 
•  An additional document with Ex safety instructions is 
available for measuring systems that are used in potentially 
explosive atmospheres. 
•  Ex safety instructions are an integral part of this manual. As 
a result, it is crucial that the installation guidelines and 
connection values it lists are also observed.  
The icon on the name plate indicates
the following: 
 
 
 
 
 
 
Installation conditions 
 
 
General 
The following points must be observed during installation: 
•  The flow direction must correspond to the marking, if 
present 
•  The maximum torque for all flange screws must be 
complied with 
•  Secure flange screws and nuts against pipe vibration. 
•  The devices must be installed without mechanical tension 
(torsion, bending) 
•  Install flange devices / wafer-type devices with plane 
parallel counterflanges and use appropriate gaskets only 
•  Use gaskets made from a material that is compatible 
with the measuring medium and measuring medium 
temperature. 
•  Gaskets must not extend into the flow area, since 
possible turbulence could influence the accuracy of the 
device 
•  The piping may not exert any inadmissible forces or 
torques on the device. 
•  Make sure that the temperature limits are not up-scaled 
during operation of the device. 
•  Vacuum shocks in the piping should be avoided to 
prevent damage to the liners (PTFE liner). Vacuum shocks 
can destroy the device. 
•  Do not remove the sealing plugs in the cable glands until 
you are ready to install the electrical cable 
•  Make sure the gaskets for the housing cover are seated 
correctly. Carefully seal the cover. Tighten the cover 
fittings 
•  The transmitter with a remote mount design must be 
installed at a largely vibration-free location 
•  Do not expose the transmitter and sensor to direct 
sunlight. Provide appropriate sun protection as necessary 
If necessary, provide a suited means of sun protection. 
•  When installing the transmitter in a control cabinet, make 
sure adequate cooling is provided 
 
 
Devices with extended diagnostic functions 
 
For devices with extended diagnostic functions different 
installation conditions may be valid. 
For additional information, see Extended diagnostic functions 
on page 118.