18  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
 
… 5  Installation 
… Installation conditions 
 
Grounding 
 
The flowmeter sensor must be connected to ground potential. 
For technical reasons, this potential must be identical to the 
potential of the measuring medium. 
In piping made of plastic or with insulating liner, grounding of 
the measuring medium is done by installing grounding plates.  
If stray potential is present in the piping, adding a grounding 
plate on both ends of the flowmeter sensor is recommended. 
 
 
Sensor insulation 
 
 
1 Insulation 
 
Figure 15:  Insulation of the flowmeter sensor 
 
In the high temperature design, the flowmeter sensor can be 
completely thermally insulated. After the unit is installed, the 
piping and sensor must be insulated in accordance with the 
figure. 
 
 
 
 
 
 
Inlet and outlet sections 
 
 
1 Double manifold  2 Turn-off device 
Figure 16:  Inlet and outlet section, turn-off devices 
 
The measuring principle is independent of the flow profile as 
long as standing eddies do not extend into the measured value 
formation, such as may for example occur after double 
manifolds, in the event of tangential inflow, or where half-open 
gate valves are located upstream of the sensor. In such cases, 
measures must be put in place to normalize the flow profile. 
 
A Do not install fittings, manifolds, valves, etc., right before the 
flowmeter sensor. 
B Inlet / outlet sections: length of the straight piping upstream 
and downstream on the sensor. 
  Experience has shown that, in most installations, straight 
inlet sections 3 × DN long and straight outlet sections 2 × DN 
long are sufficient (DN = nominal diameter of the flowmeter 
sensor). 
  For test stands, the reference conditions of 10 × DN straight 
inlet and 5 × DN straight outlet must be provided, in 
accordance with EN 29104 / ISO 9104. 
  Valves or other turn-off devices should be installed in the 
outlet section. 
  Valve flaps must be installed so that the valve damper plate 
does not extend into the flowmeter sensor.