120  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
 
… 9  Diagnosis / error messages 
… Extended diagnostic functions 
 
Monitoring the electrode impedance 
 
The measurement monitors the impedance between the 
measuring electrode and grounding and activates an alarm if the 
impedance drops below a limit. The function is activated 
together with the conductivity measurement. 
 
Requirements for use: 
•  Conductivity of the measuring medium: 
20 to 20000 μS/cm. 
 
Additional installation conditions: 
•  When using plastic piping, install a grounding plate at the 
front and back of the device. 
•  There must not be any deposits on the measuring 
electrodes. 
•  The measuring tube must always be completely full, and 
the measuring medium must have only minor deviations 
in conductivity. 
 
 
 
 
 
 
Measurements on the flowmeter 
 
Coil inductance, coil current, coil resistance 
The diagnosis of the coil in the sensor includes coil inductance, 
current and resistance. 
 
Flowmeter sensor temperature 
The coil temperature monitoring triggers an alarm via the digital 
output, if configured. 
The minimum and maximum alarm value for the coil temperature 
can be set. 
The coil temperature is a function of the ambient temperature 
and measuring medium temperature. 
Compliance with the temperature specification of the sensor 
liner can thus be monitored. 
 
 
Setup 
 
Menu / parameter  Description 
Diagnostics / ...Diagnosis Control / ...Diagnosis Coil 
Coil Diag On/Off  Activate the function. 
Coil Resistor  Display the coil resistance. 
Coil Current  Display the coil current. 
Coil Inductance  Display the coil inductance. 
Coil Temperature  Display the coil temperature within the sensor. 
Coil Temperature Adj  Measurement of coil temperature must be set in 
accordance with the conditions on-site. 
Temperature measured with a separate 
thermometer can be entered here. 
Coil Temp. Min Alarm  Min. and max. alarm for the sensor temperature 
(coil temperature). Can be used to monitor the 
temperature limit of the meter tube liner 
Coil Temp. Max Alarm 
 
 
Transmitter monitoring 
 
Monitoring the temperature of the electronic unit in the 
transmitter triggers an alarm via the digital output, if 
configured. 
In the ‘...Diagnosis Values’, the current temperature as well as the 
smallest and largest previously measured temperature is 
displayed.